DOI 10.3760/cma.j.issn.0376-2491.2016.14.014
[A new device for removing Shang ring after circumcision ( a report of 257 cases)].
Аннотация
OBJECTIVE
To evaluate the clinical value of a new device designed by ourselves for removing the disposable circumcision device (Shang ring).
METHODS
There were 264 patients undergoing Shang ring circumcision from August 2013 to June 2015. A total of 7 cases were lost follow-up and were excluded. The new device was applied for removing Shang ring in 150 patients after operation in Experimental group, and that was not applied in 107 cases in Control group. Values of basic demographic data, time for removing Shang ring, pain score, scar satisfaction, surgical satisfaction and complications were recorded and compared.
RESULTS
Time for removing Shang ring and pain score in Experimental group and Control group was 61.8±9.9 vs 127.4±41.6 s, 4.1±1.3 vs 6.8±1.4, respectively (both P<0.05). Satisfaction of surgery, percentage of willing to recommend Shang ring surgery to others in Experiment group and Control group was 9.3±0.7 vs 7.5±1.3, and 95.3% vs 79.4%, respectively (both P<0.05). Surgical satisfaction was correlated with pain score, time of removing Shang ring and appearance of incision scar (P<0.05).
CONCLUSION
The new device for removing the Shang ring could remove the ring rapidly, alleviate pain and improve surgical satisfaction.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.