m+Medelevator.RU
← Поиск публикаций
Child's nervous system : ChNS : official journal of the International Society for Pediatric Neurosurgery · 2022

Proximal ventricular shunt catheter occlusion model.

СтатьяНазвание и аннотация: Перевод в очереди
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Qi David, Patel Anup, Dunwoody Robert, McCall Shannon, Bach Sarah, Lin Julian

Аннотация

PURPOSE: Proximal ventricular shunt catheter occlusion remains a problematic cause of shunt malfunction, and there is no consistent in vivo or in vitro model to help clinicians and researchers study this phenomenon.

METHODS: An in vitro model utilizing standard proximal ventricular catheter and biological occluding agents mimicking choroid plexus was designed, constructed, and calibrated to occlude consistently within a specified timeframe. Hydrostatic pressure differential of 100 cmH2O was used as a driving force to generate flow through the catheter. Chalaza and vitelline membranes were harvested from avian eggs and used as occluding agents. Successful occlusion was defined as a greater than 90% reduction in volumetric flow rate through distal outlet. Histological sections of occluded catheters were performed and interpreted by a neuropathologist.

RESULTS: Initial trials demonstrated successful standard catheter occlusion within 24 h using chalaza, vitelline membrane, and combination treatments. Repeat trials demonstrated consistency in successful occlusion within 5 min utilizing only vitelline membrane treatment. Histopathology demonstrated the vitelline membrane to consist of a thin, superficial layer of extraembryonic ectoderm; the chalaza was observed to consist of strands of mucin protein.

CONCLUSIONS: An in vitro model of proximal ventricular shunt catheter occlusion was developed and calibrated for successful occlusion within 5 min. Future studies may utilize this model to rapidly test occlusion-resistant shunt designs and de-obstruction techniques.