DOI 10.1016/j.athoracsur.2016.03.082
Complete Transversal Disc Fracture in a Björk-Shiley Delrin Mitral Valve Prosthesis 43 Years After Implantation.
Аннотация
A patient who underwent previous implantation of a mitral valve replacement with a Björk-Shiley Delrin (BSD) mitral valve prosthesis during infancy was admitted to our institution 43 years later after an episode of syncope and cardiac arrest. Under extreme hemodynamic instability, a mitral valve prosthetic dysfunction causing massive mitral regurgitation was identified. The patient underwent an emergent cardiac operation, and a complete disc fracture with partial disc migration was found. Exceptional cases of mechanical prosthetic heart valve fracture exist. We report the first case of complete transversal disc rupture of a BSD mitral valve prosthesis after the longest period of implantation ever reported in that position.