Songy Bernard, Guernou Mohamed, Lussato David, Queneau Mathieu, Bonardel Gerald, Grellier Jean-François, Flahaut Ludovic, Nguyen Duc

СтатьяПеревод в очереди

Аннотация

BACKGROUND

We studied the impact of technetium-99m (99mTc) in the thallium-201 (201Tl) energy window (70 keV) to determine if CZT cardiac cameras allow us to perform simultaneous dual-isotope acquisition for myocardial perfusion imaging.

METHODS

We included 117 consecutive patients. We injected 0.7 MBq/kg of 201Tl at stress, performed the first scan (image T1), then injected at rest 2 MBq/kg of 99mTc-tetrofosmin and immediately acquired a second scan with reconstruction in the energy window of thallium (image T2). A corrected thallium image was created by the subtraction of 99mTc downscattered photons (image TS). We compared spectra, image quality, and semiquantitative scores on T1, T2, and TS images.

RESULTS

Though T2 images were of worse quality, TS images were of equal quality compared to T1 images in most cases. Scores show an underestimation of abnormalities in 20% of patients on T2 images and in 10% on TS images.

CONCLUSIONS

Despite the improved energy resolution of CZT cameras, downscatter of technetium in the 201Tl window leads to an underestimation of the pathological territory in 10% to 20% of cases. It does not allow us to use simultaneous dual-isotope acquisition in clinical practice without additional tools for scatter correction.

Для цитирования: Songy Bernard, Guernou Mohamed, Lussato David, Queneau Mathieu, Bonardel Gerald, Grellier Jean-François, Flahaut Ludovic, Nguyen Duc. Feasibility of simultaneous dual isotope acquisition for myocardial perfusion imaging with a cadmium zinc telluride camera.. Journal of nuclear cardiology : official publication of the American Society of Nuclear Cardiology. 2018. DOI: 10.1007/s12350-018-1452-z.