Self-monitoring of intraocular pressure using Icare HOME tonometry in clinical practice.

Murabito Alessandra, Mergiotti Marco, Capurro Valeria, Loffreda Alessia, Li Mingchuan, Peretto Paola, Ren Kai, Raimondi Andrea, Tacchetti Carlo, Diviani Dario, Pedemonte Nicoletta, Hirsch Emilio, Ghigo Alessandra

СтатьяПеревод пока не подготовлен

Аннотация

Purpose: To determine the value of self-monitoring of diurnal intraocular pressure (IOP) by Icare Home rebound tonometer in patients with glaucoma and ocular hypertension.

Methods

Patients with open-angle glaucoma or ocular hypertension, controlled IOP at office visits, and at least 3 years of follow-up in the glaucoma clinic were included. Progression of glaucoma was based on medical records and defined by documented structural and/or visual field change. Patients were trained to correctly perform self-tonometry and instructed to measure diurnal IOP in a home setting for 3 days. IOP characteristics (mean, peak IOP, fluctuation of IOP as range, and SD of IOP) were documented and compared between the progressive and stable eyes.

Results

Ninety-four patients (50 females) with a mean (SD) age of 57.1 (14.7) years were included. Among the 94 eyes from 94 subjects, 72 (76.6%) eyes had primary open-angle glaucoma, ten (10.6%) had pigmentary glaucoma, four (4.3%) had exfoliative glaucoma, and eight (8.5%) eyes had ocular hypertension. Thirty-six eyes showed progression and 58 eyes were stable. Patients with progression were older than those with stable disease (mean (SD) 65.8 (8.4) years vs 51.7 (15.3) years, P<0.001). The progression group had higher average IOP (mean (SD) 15.8 (4.0) mmHg vs 13.3 (3.7) mmHg, P=0.002), peak IOP (mean (SD) 21.8 (5.8) mmHg vs 18.6 (4.8) mmHg, P=0.01), and greater IOP fluctuation range (mean (SD) 11.6 (4.8) vs 9.1 (3.5) mmHg, P=0.011) compared to non-progression group.

Conclusion

Self-monitoring of IOP using Icare Home tonometry provides more complete data on variability of IOP to assist in the management of glaucoma.

Для цитирования: Murabito Alessandra, Mergiotti Marco, Capurro Valeria, Loffreda Alessia, Li Mingchuan, Peretto Paola, Ren Kai, Raimondi Andrea, Tacchetti Carlo, Diviani Dario, Pedemonte Nicoletta, Hirsch Emilio, Ghigo Alessandra. Self-monitoring of intraocular pressure using Icare HOME tonometry in clinical practice.. Clinical ophthalmology (Auckland, N.Z.). 2019. DOI: 10.2147/opth.s198846.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.