Are locking plates better than non-locking plates for treating distal tibial fractures?

Bastias Christian, Henríquez Hugo, Pellegrini Manuel, Rammelt Stefan, Cuchacovich Natalio, Lagos Leonardo, Carcuro Giovanni

СтатьяПеревод в очереди

Аннотация

BACKGROUND

Locking and non-locking plates has been used for distal tibia fracture osteosynthesis. Sufficient evidence to favor one implant over the other is lacking in the current literature. Our aim is to compare them in terms of fracture healing, alignment, functional outcome, complications.

METHODS

Sixty-eight patients operated on using a percutaneous plate were retrospectively reviewed. They were divided into two groups: in group 1 (28 patients) a 4.5mm narrow conventional dynamic compression plate (DCP) was used. In group 2 (40 patients) a titanium locked compression plate (LCP) was used.

RESULTS

Mean time to union was 16.2 and 15.4 weeks for group 1 and 2, respectively (p=0.618). 11 patients (39.3%) in group 1 and 4 patients (10%) in group 2 showed malalignment (p=0.016). AOFAS scores at follow up were 89 and 88 in groups 1 and 2, respectively. Implant removal was necessary in 9 cases (32.1%) and 4 cases (10%) in group 1 and group 2, respectively (p=0.042). Three patients (10.7%) in group 1 and three patients (7.5%) in group 2 had an infection.

CONCLUSIONS

Both plating systems have similar results in terms of time to union, infection, and AOFAS scores. The LCP seems superior with respect to alignment and the need for implant removal.

Для цитирования: Bastias Christian, Henríquez Hugo, Pellegrini Manuel, Rammelt Stefan, Cuchacovich Natalio, Lagos Leonardo, Carcuro Giovanni. Are locking plates better than non-locking plates for treating distal tibial fractures?. Foot and ankle surgery : official journal of the European Society of Foot and Ankle Surgeons. 2014. DOI: 10.1016/j.fas.2013.12.004.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.