m+Medelevator.RU
← Поиск публикаций
Clinical chemistry and laboratory medicine · 2025

An isotope dilution-liquid chromatography-tandem mass spectrometry-based candidate reference measurement procedure for the quantification of cortisone in human serum and plasma.

СтатьяПоддержка исследования: вне правительства СШАНазвание и аннотация: Перевод в очереди
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Ott Myriam, Singh Neeraj, Kubicova Marie, Bauland Friederike, Köppl Daniel, Gaudl Alexander, Geistanger Andrea, Ceglarek Uta, Rauh Manfred, Geletneky Christian, Taibon Judith

Аннотация

OBJECTIVES: Cortisone is an inert precursor/metabolite of the potent steroid hormone cortisol. Measurement of serum cortisone levels and the cortisol-cortisone ratio can be useful for the diagnosis of dysfunction in the regulation of cortisol levels (i.e., severe and subtle apparent mineralocorticoid excess, low-renin primary aldosteronism). Therefore, an isotope dilution-liquid chromatography-tandem mass spectrometry (ID-LC MS/MS)-based candidate reference measurement procedure (RMP) to quantify cortisone in human serum/plasma was developed and validated.

METHODS: Quantitative nuclear magnetic resonance (qNMR) was utilized to assign absolute content (g/g) and SI-traceability to reference materials used as primary calibrators. A supported liquid extraction sample preparation protocol as well as a two-dimensional heart-cut LC approach for LC-MS/MS analysis were employed to mitigate matrix effects and prevent co-elution of interferences. Selectivity was determined by analyzing a matrix sample containing the analyte, the internal standard and six potential interferents. A post-column infusion experiment and a comparison of standard line slopes were performed to evaluate matrix effects. An extensive protocol over five days was applied to determine precision, accuracy and trueness. Measurement uncertainty (MU) was evaluated in compliance with current guidelines.

RESULTS: This RMP is suitable for analyzing cortisone within the 0.0800-120 ng/mL (0.222-333 nmol/L) range, demonstrating selectivity, sensitivity and matrix independence. Intermediate precision was ≤3.4 %, repeatability was ≤2.9 % across all concentration levels and relative mean bias ranged from -3.7 to 2.8 % across all tested matrices and concentrations. Expanded MU (k=2) for target value assignment (n=6) ranged from 2.1 to 5.5 %, irrespective of concentration or sample type.

CONCLUSIONS: This RMP allows for accurate and reproducible determination of cortisone in human serum and plasma. Implementation of this method supports routine assay standardization and patient sample measurement with confirmed traceability.

Предметные рубрики MeSH

ИП Урсолов Владлен Сергеевич

ИНН 500118482352 · ОГРНИП 314500117700049

Реквизиты и контакты