Geometric Analysis to Determine Kinking and Shortening of Bridging Stents After Branched Endovascular Aortic Repair.

Finotello Alice, Spinella Giovanni, Notini Giulia, Palombo Domenico, Pratesi Giovanni, Mambrini Simone, Auricchio Ferdinando, Conti Michele, Pane Bianca

СтатьяПеревод в очереди

Аннотация

PURPOSE: To evaluate bridging stent geometry in patients who underwent branched endovascular aortic repair (B-EVAR) and to correlate the outcomes with intrinsic bridging stent characteristics aiming to identify the stent(s) that guarantees the best performance.

METHODS

Pre-operative and post-operative computed tomography images of all patients undergoing B-EVAR between September 2016 and April 2019 were retrospectively analyzed. Following geometrical features were measured: target vessel take-off angle (TOA); longitudinal stent shortening; shape index (SI), intended as ratio between minimum and maximum diameter of the lumen cross sections, averaged on three segments: zone 1 (proximal stented zone), zone 2 (intermediate), and zone 3 (distal).

RESULTS

Thirty-eight branches (8 right (RRA) and 8 left renal arteries (LRA), 11 superior mesenteric arteries (SMA), 11 celiac trunks (CTR)) were treated. Fluency (Bard Peripheral Vascular), COVERA (Bard Peripheral Vascular), and VBX (WLGore&Assoc) stent-grafts were implanted in 10, 12, and 16 branches, respectively. Pre-operative TOA was more acute in RRA and LRA when compared to CTR and SMA, and straightened in post-operative configuration (109.86 ± 28.65° to 150.27 ± 21.0°; P < 0.001). Comparable values of SI among the stent types were found in zone 1 (P = 0.08), whereas higher SI in VBX group was detected in zones 2 (P < 0.001) and 3 (P < 0.001). The VBX group was also the most affected by stent shortening (11.12 ± 5.65%; P = 0.001).

CONCLUSION

Our early experience showed that the VBX stent offers greater stent circularity than the other devices even if a greater shortening has been observed drawing attention with regards to the decision of the nominal stent length.

Для цитирования: Finotello Alice, Spinella Giovanni, Notini Giulia, Palombo Domenico, Pratesi Giovanni, Mambrini Simone, Auricchio Ferdinando, Conti Michele, Pane Bianca. Geometric Analysis to Determine Kinking and Shortening of Bridging Stents After Branched Endovascular Aortic Repair.. Cardiovascular and interventional radiology. 2021. DOI: 10.1007/s00270-021-02773-w.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.