Accuracy of home blood pressure readings: monitors and operators.
Аннотация
OBJECTIVE
To evaluate the accuracy of automated digital blood pressure monitoring devices and operators in the community. Also, we tested the effects of a simple education program, and looked for arm-arm differences.
DESIGN
Subjects who had bought their own automated digital blood pressure monitor were recruited via an advertisement in the local newspaper. On arrival, they were asked to record their blood pressure exactly as they would at home. The investigator noted any technique deficiencies then corrected them. Blood pressures were then recorded by the investigator and the subject, on opposite arms, simultaneously, and repeated with the arms switched. Finally, subjects recorded their blood pressure again. The subjects' readings were compared to the average of monitor and mercury readings using Bland-Altman methods.
RESULTS
A total of 80 subjects were tested. Before educating, subjects' systolic blood pressure (SBP) readings were +5.8+/-6.4 (standard deviation) mmHg greater than the mean of all readings, and diastolic blood pressure (DBP) were +1.3+/-4.0 mmHg; after educating they were +1.3+/-4.0 and -1.3+/-2.7 respectively. The monitors, as a group, were accurate, and met British Hypertension Society and AAMI highest standards. We found no differences among monitors that had been validated (n=26) and those that had not. There were differences between the arms: 5.3+/-5.2 mmHg for SBP and 3.4+/-3.3 mmHg for DBP. Most patients had never been informed by anyone of proper blood pressure measuring techniques.
CONCLUSIONS
We conclude that home blood pressure measurement, as practiced in our community, is prone to error, mostly due to mistakes by the operator. These can easily be corrected, so that readings become more accurate. Attention should be paid to arm-arm differences.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.