Transcutaneous bilirubinometry in very low birthweight infants.
Аннотация
AIM: To evaluate whether transcutaneous bilirubinometry (TcB) would be a reliable and efficient screening technique for hyperbilirubinaemia in very low birthweight (VLBW, < or =1500 g) infants in an intensive care unit setting.
METHODS
TcB measurements (Minolta Airshield Jaundice Meter JM-102, Osaka, Japan) were obtained immediately before or within 10 min following routine blood sampling for plasma bilirubin concentration measurements in 124 VLBW infants not receiving phototherapy. The relationship between the two techniques was analysed by linear regression analysis. A plasma bilirubin > or =150 micromol/l was defined as hyperbilirubinaemia. The sensitivity and specificity of possible TcB cut-off readings to detect hyperbilirubinaemia was evaluated.
RESULTS
There was a significant correlation between the measurements of both techniques (p < 0.0001, r = 0.68). In the present study, a TcB cut-off reading of 14 would have reduced the need for plasma bilirubin measurements by 26% without missing true hyperbilirubinaemia.
CONCLUSION
The data suggest that TcB will improve VLBW infant care in an intensive care unit setting by reducing the need for invasive bilirubin concentration measurements.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.