Comparing prostate diffusion weighted images reconstructed with a commercial deep-learning product to a deep learning phase corrected model at 1.5 T.

Cochran Rory L, Bradley William R, Dhami Ranjodh S, Milshteyn Eugene, Pohl Maximilian, Ghosh Soumyadeep, Nakrour Nabih, Lan Patricia, Wang Xinzeng, Guidon Arnaud, Harisinghani Mukesh G

СтатьяСравнительное исследованиеПеревод в очереди

Аннотация

PURPOSE: To determine whether a new deep learning (DL) based phase corrected (DLPC) reconstruction model can enhance image quality of diffusion weighted images of the prostate acquired at 1.5 T compared to a commercially available DL based product.

METHODS AND MATERIALS: A retrospective study of 30 consecutive patients undergoing conventional multiparametric MRI (mpMRI) of the prostate on a single 1.5 T scanner was performed. Diffusion image datasets reconstructed with a commercially available DL product and a new DLPC model were assessed. Qualitative image assessment was performed by three board certified radiologists using a 5-point Likert scale across four features and inter-rater agreement was estimated using Gwet's AC2 statistic. Quantitative image comparison was performed by assessing SNR of acquired intermediate b-value (b = 1000 s/mm2) diffusion images. The Wilcoxon matched-pairs signed rank test was used to assess differences between techniques. Image noise was assessed using the edge function.

RESULTS

Median patient age was 70 years (interquartile range: 66.0-75.3). All radiologists perceived less noise and better image quality for all DLPC image sets compared to commercial DL images (p < 0.05). Significantly higher SNR was observed for the acquired intermediate b-value diffusion images reconstructed with DLPC (median SNR: 49.4 vs 27.5; p < 0.001), and mean ADC values did not significantly differ between DLPC and DL images (p = 0.63). Edge analyses demonstrated significantly reduced noise for DLPC images (p < 0.001).

CONCLUSIONS

DLPC image reconstruction of diffusion weighted prostate image datasets reduces image noise and improves SNR over a commercial DL product at 1.5 T.

Для цитирования: Cochran Rory L, Bradley William R, Dhami Ranjodh S, Milshteyn Eugene, Pohl Maximilian, Ghosh Soumyadeep, Nakrour Nabih, Lan Patricia, Wang Xinzeng, Guidon Arnaud, Harisinghani Mukesh G. Comparing prostate diffusion weighted images reconstructed with a commercial deep-learning product to a deep learning phase corrected model at 1.5 T.. Clinical imaging. 2025. DOI: 10.1016/j.clinimag.2025.110681.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.