DOI 10.1007/s00167-018-4976-5

Медицинская библиотека · Knee surgery, sports traumatology, arthroscopy : official journal of the ESSKA · 2018

It is safe and effective to use all inside meniscal repair devices for posteromedial meniscal 'ramp' lesions.

Heilpern Giles, Stephen Jo, Ball Simon, Amis Andrew, Williams Andy

Оценочное исследованиеСтатьяПеревод в очереди

Аннотация

PURPOSE: Recently, it has been recognized that meniscocapsular ('ramp') lesions of the posterior one-third of the medial meniscus frequently occur during injuries causing ACL rupture, and that these lesions are easily missed at arthroscopy. Furthermore, it is clear that these lesions are biomechanically significant, adding to the deficits caused by ACL rupture, and that their repair can reverse this. The efficacy of an all inside repair technique has been questioned by some authors and by those who advocate a suture shuttle technique via an accessory posteromedial portal. The use of Ultra FastFix and FastFix 360 meniscal repair devices to repair posteromedial meniscocapsular separations was investigated in terms of safe deployment and the effectiveness.

METHODS

Twenty cadaveric fresh frozen knees were used-ten in each of two groups. A ramp lesion was created using a Beaver knife. The lesion was then repaired with either 4 Ultra FastFix (Smith and Nephew) or 4 FastFix 360 (Smith and Nephew) meniscal repair devices. The knees were put through a standardized loading cycle consisting of 10 Lachman's tests and ten maximum loading manual anterior drawer tests at 90° of flexion. Each knee was then flexed and extended fully ten times. The specimens were sectioned just proximal to the menisci and each suture anchor identified and its position recorded and photographed.

RESULTS

In the Ultra FastFix group, a single anchor was found to be in an intra-articular position-a failure rate of 2.5%. In the FastFix 360 group, 5 anchors failed-a 12.5% failure rate. In all cases, the anchors were attached to their suture and so not truly loose within the joint.

CONCLUSIONS

This study confirms the safe and effective deployment of an all inside repair device for repair of medial meniscal 'ramp' lesions, and therefore its use is advocated in treating these difficult lesions. Ultra FastFix had the lower failure rate of 2.5%, which the authors believe is acceptable, and makes this device preferable to the FastFix360.

Для цитирования: Heilpern Giles, Stephen Jo, Ball Simon, Amis Andrew, Williams Andy. It is safe and effective to use all inside meniscal repair devices for posteromedial meniscal 'ramp' lesions.. Knee surgery, sports traumatology, arthroscopy : official journal of the ESSKA. 2018. DOI: 10.1007/s00167-018-4976-5.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.