CT of facial fracture fixation: an experimental study of artefact reducing methods.

Peltola Elina M, Mäkelä Teemu, Haapamäki Ville, Suomalainen Anni, Leikola Junnu, Koskinen Seppo K, Kortesniemi Mika, Koivikko Mika P

СтатьяПеревод в очереди

Аннотация

OBJECTIVES

This study aimed to determine the optimal post-operative CT imaging method that enables best visualization of facial bony structures in the vicinity of osteosynthesis material.

METHODS

Conducted at Töölö Hospital (Helsinki, Finland), this study relied on scanning a phantom with CBCT, 64-slice CT and high-definition multislice CT with dual-energy scan (providing monochromatic images of 70-, 100-, 120- and 140-keV energy levels) and iterative reconstruction (IR) methods. Two radiologists assessed the image quality, and the assessments were analyzed. In addition, a physicist performed a semi-quantitative analysis of the metal-induced artefacts.

RESULTS

The three subjects most easily assessed were the loose screw and both the bone structure and the fracture further away from the screw and the plate. Soft tissues adjacent to the screw and the plate remained more difficult for assessment. Both image interpreters agreed that the artefacts disturbed their assessments under dual energy. Metal artefacts disturbed the least under multislice CT with IR [adaptive statistical iterative reconstruction (ASiR) and VEO]. Neither interpreter found metal suppression helpful in CBCT.

CONCLUSIONS

CBCT with or without a metal artefact reduction algorithm was not optimal for post-operative facial imaging compared with multislice CT with IR. Multislice CT with ASiR filtering offered good image quality performance with fast image volume reconstruction, representing the current sweet spot in post-operative maxillofacial imaging.

Для цитирования: Peltola Elina M, Mäkelä Teemu, Haapamäki Ville, Suomalainen Anni, Leikola Junnu, Koskinen Seppo K, Kortesniemi Mika, Koivikko Mika P. CT of facial fracture fixation: an experimental study of artefact reducing methods.. Dento maxillo facial radiology. 2016. DOI: 10.1259/dmfr.20160261.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.