In vitro elimination of anti-infective drugs by the Seraph® 100 Microbind® affinity blood filter.
Аннотация
BACKGROUND
In August 2019, the European Union licensed the first ever haemoperfusion device aimed to reduce pathogens in the blood. The core of the adsorber consists of ultra-high molecular weight polyethylene beads with endpoint-attached heparin. These beads utilize pathogen inherent adhesion mechanisms to reduce pathogen load. So far, it is unknown whether the device has an effect on anti-infective drug concentrations. The aim of this study was to investigate the in vitro adsorption of multiple anti-infective drugs from human plasma.
METHODS
In this in vitro study, 18 anti-infective drugs were administered to human donor plasma and pumped through the heparin-coated pathogen adsorber (Seraph® 100 Microbind®Affinity Blood Filter; ExThera Medical Corp., Martinez, CA, USA) at a plasma flow rate of 250 mL/min for 60 min. Pre- and post-adsorber plasma samples were quantified after 5, 15, 30 and 60 min.
RESULTS
We found a reduction ratio (RR) in anti-infective plasma levels between -1% and 62%. This decrease occurred mainly in the first 5 min of the experiment (RR0-5 -4 to 62%). Mean plasma clearance rates ranged between -11.93 mL/min (fluconazole) and 4.86 mL/min (clindamycin). The highest RRs were measured for aminoglycosides (tobramycin 62% and gentamycin 59%).
CONCLUSIONS
The elimination of anti-infective drugs by the Seraph is neglectable in all but 2 of 18 of the investigated substances. Aminoglycosides may be adsorbed by the device during their first pass.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.