Pseudoatrial flutter: artifact simulating atrial flutter caused by a transcutaneous electrical nerve stimulator (TENS).

Weitz S H, Tunick P A, McElhinney L, Mitchell T, Kronzon I

Клинический случайСтатьяПеревод в очереди

Аннотация

A patient with previous actual atrial flutter had what appeared to be atrial flutter seen on Holter monitoring during treatment with a transcutaneous electrical nerve stimulation (TENS) unit. This proved to be artifact caused by the unit rather than a true arrhythmia. The artifact was reproduced in a normal volunteer by application of a TENS unit.

Для цитирования: Weitz S H, Tunick P A, McElhinney L, Mitchell T, Kronzon I. Pseudoatrial flutter: artifact simulating atrial flutter caused by a transcutaneous electrical nerve stimulator (TENS).. Pacing and clinical electrophysiology : PACE. 1997. DOI: 10.1111/j.1540-8159.1997.tb05478.x.