The Journal of thoracic and cardiovascular surgery · 2018
Prospective US investigational device exemption trial of a sutureless aortic bioprosthesis: One-year outcomes.
Клиническое исследованиеСтатьяМногоцентровое исследованиеVideo-Audio MediaНазвание и аннотация: Перевод в очереди
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Suri Rakesh M, Javadikasgari Hoda, Heimansohn David A, Weissman Neil J, Ailawadi Gorav, Ad Niv, Aldea Gabriel S, Thourani Vinod H, Szeto Wilson Y, Michler Robert E, Michelena Hector I, Dabir Reza, Fontana Gregory P, Kessler William F, Moront Michael G, Brunsting Louis A, Griffith Bartley P, Montoya Alvaro, Subramanian Sreekumar, Mostovych Mark A, Roselli Eric E
Аннотация
OBJECTIVES: We performed a prospective, single-arm clinical trial approved under a Food and Drug Administration Investigational Device Exemption to assess safety and efficacy of Perceval, a sutureless bovine pericardial aortic valve representing the initial US experience.
METHODS: From June 2013 to January 2015, 300 patients (mean age 76.7 ± 7.7 years, 54.3% men, 37.3% ≥80 years, median Society of Thoracic Surgeons Predicted Risk of Mortality 2.8%), underwent Perceval valve implantation at 18 centers across the United States. Twenty patients (6.7%) had a bicuspid aortic valve and 5 (1.7%) patients had previous aortic valve replacement. A minimally invasive approach was used in 80 (26.7%) and concomitant procedures were performed in 113 (37.8%) patients.
RESULTS: Two hundred eighty-nine patients (96.3%) were successfully implanted. Operative mortality (≤30 days) was 1.3% (n = 4) (observed to expected ratio of 0.40). One-year results included all-cause mortality in 5.2% (n = 15), stroke in 1% (n = 3), and endocarditis in 1.7% (n = 5). New permanent periprocedural pacemaker rate was 10.7% (n = 30/281); 2.5% (n = 7/281) resulted from third-degree atrioventricular block. One-year valve-related reoperation was 2.1% (n = 6). At 1-year follow-up, 98% of patients were in New York Heart Association class I/II, left ventricular mass index decreased from 103.5 ± 30.1 g/m2 at discharge to 95.8 ± 27.1 g/m2 (P = .001), and 3 (1.3%) moderate paravalvular leaks were identified. Health-related quality of life score increased from 62.7 ± 21.8 before surgery to 85.5 ± 17.8 at 1 year (P < .001).
CONCLUSIONS: These results confirm the safety and effectiveness of the Perceval sutureless aortic valve replacement in study patients with lower mortality than expected from a risk prediction model. Persistent hemodynamic benefit and improvement in quality of life at 1 year support the importance of this device in the management of aortic valve disease.
METHODS: From June 2013 to January 2015, 300 patients (mean age 76.7 ± 7.7 years, 54.3% men, 37.3% ≥80 years, median Society of Thoracic Surgeons Predicted Risk of Mortality 2.8%), underwent Perceval valve implantation at 18 centers across the United States. Twenty patients (6.7%) had a bicuspid aortic valve and 5 (1.7%) patients had previous aortic valve replacement. A minimally invasive approach was used in 80 (26.7%) and concomitant procedures were performed in 113 (37.8%) patients.
RESULTS: Two hundred eighty-nine patients (96.3%) were successfully implanted. Operative mortality (≤30 days) was 1.3% (n = 4) (observed to expected ratio of 0.40). One-year results included all-cause mortality in 5.2% (n = 15), stroke in 1% (n = 3), and endocarditis in 1.7% (n = 5). New permanent periprocedural pacemaker rate was 10.7% (n = 30/281); 2.5% (n = 7/281) resulted from third-degree atrioventricular block. One-year valve-related reoperation was 2.1% (n = 6). At 1-year follow-up, 98% of patients were in New York Heart Association class I/II, left ventricular mass index decreased from 103.5 ± 30.1 g/m2 at discharge to 95.8 ± 27.1 g/m2 (P = .001), and 3 (1.3%) moderate paravalvular leaks were identified. Health-related quality of life score increased from 62.7 ± 21.8 before surgery to 85.5 ± 17.8 at 1 year (P < .001).
CONCLUSIONS: These results confirm the safety and effectiveness of the Perceval sutureless aortic valve replacement in study patients with lower mortality than expected from a risk prediction model. Persistent hemodynamic benefit and improvement in quality of life at 1 year support the importance of this device in the management of aortic valve disease.
Предметные рубрики MeSH
AgedAged, 80 and overAortic ValveBicuspid Aortic Valve DiseaseBioprosthesisFemaleHeart Valve DiseasesHeart Valve ProsthesisHeart Valve Prosthesis ImplantationHemodynamicsHumansMalePostoperative ComplicationsProspective StudiesProsthesis DesignRecovery of FunctionRisk FactorsSutureless Surgical ProceduresTime FactorsTreatment OutcomeUnited States