Annals of vascular surgery · 2021
Defining Duplex Ultrasound Criteria for In-Stent Restenosis of the Superior Mesenteric Artery.
СтатьяНазвание и аннотация: Перевод в очереди
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Green Jessica, Ryer Evan, Borden Nicholas, Ali Bilal, Garvin Robert, Yang Andrew, Hashmi Ammar, Salzler Gregory, Elmore James
Аннотация
BACKGROUND: This study sought to define duplex ultrasound (DUS) velocity criteria predicting ≥70% stenosis in superior mesenteric artery (SMA) stents by correlating in-stent peak systolic velocity (PSV) with computed tomographic angiography (CTA) measurements of percent stenosis.
METHODS: A retrospective review of 109 patients undergoing SMA stenting between 2003 and 2018 was conducted at a single institution. Thirty-seven surveillance duplex ultrasound studies were found to have a CTA performed within 30 days of study completion. Bare metal (n = 20) and covered stents (n = 17) were included. Velocities were paired to in-stent restenosis (ISR) measured by mean vessel diameter reduction on SMA centerline reconstructions from CTA. Receiver operating characteristic (ROC) curves was generated and logistic regression models for ≥70% ISR probability were used to define velocity criteria in the stented SMA.
RESULTS: At a PSV of 300 cm/sec, the sensitivity is 100% and specificity 80% for a ≥70% in-stent SMA stenosis. At a PSV of 400 cm/sec, the sensitivity and positive predictive value (PPV) is 63% and the specificity and negative predictive value (NPV) is 90%. A PSV of 450 cm/sec was consistent with the highest specificity (100%) and PPV (100%) but lower sensitivity (50%) and NPV (87.9%). One patient with a PSV of 441 cm/sec on surveillance DUS died from complications of acute-on-chronic mesenteric ischemia.
CONCLUSIONS: A PSV of 400 cm/sec on mesenteric DUS can predict ≥70% ISR with high sensitivity and should be considered as a diagnostic threshold for SMA in-stent restenosis.
METHODS: A retrospective review of 109 patients undergoing SMA stenting between 2003 and 2018 was conducted at a single institution. Thirty-seven surveillance duplex ultrasound studies were found to have a CTA performed within 30 days of study completion. Bare metal (n = 20) and covered stents (n = 17) were included. Velocities were paired to in-stent restenosis (ISR) measured by mean vessel diameter reduction on SMA centerline reconstructions from CTA. Receiver operating characteristic (ROC) curves was generated and logistic regression models for ≥70% ISR probability were used to define velocity criteria in the stented SMA.
RESULTS: At a PSV of 300 cm/sec, the sensitivity is 100% and specificity 80% for a ≥70% in-stent SMA stenosis. At a PSV of 400 cm/sec, the sensitivity and positive predictive value (PPV) is 63% and the specificity and negative predictive value (NPV) is 90%. A PSV of 450 cm/sec was consistent with the highest specificity (100%) and PPV (100%) but lower sensitivity (50%) and NPV (87.9%). One patient with a PSV of 441 cm/sec on surveillance DUS died from complications of acute-on-chronic mesenteric ischemia.
CONCLUSIONS: A PSV of 400 cm/sec on mesenteric DUS can predict ≥70% ISR with high sensitivity and should be considered as a diagnostic threshold for SMA in-stent restenosis.
Предметные рубрики MeSH
AgedAged, 80 and overBlood Flow VelocityChronic DiseaseComputed Tomography AngiographyEndovascular ProceduresFemaleHumansMaleMesenteric Artery, SuperiorMesenteric IschemiaMesenteric Vascular OcclusionMiddle AgedPredictive Value of TestsRecurrenceReproducibility of ResultsRetrospective StudiesRisk AssessmentRisk FactorsSplanchnic CirculationStentsTime FactorsTreatment OutcomeUltrasonography, Doppler, DuplexVascular Patency