DOI 10.1038/jp.2014.148

Медицинская библиотека · Journal of perinatology : official journal of the California Perinatal Association · 2014

A comparison of mid-forehead and axillary temperatures in newborn intensive care.

Robertson-Smith J, McCaffrey F T, Sayers R, Williams S, Taylor B J

Сравнительное исследованиеСтатьяПеревод в очереди

Аннотация

OBJECTIVE

To evaluate accuracy of mid-forehead (MFH) thermometry compared with digital axilla (DAT) temperatures in infants in newborn intensive care.

STUDY

DESIGN

A comparative study of MFH and DAT temperatures of newborn infants receiving tertiary-level intensive care. All admissions were considered and the following exclusion criteria applied: 'in extremis', hypoxic ischemic encephalopathy or non-English-speaking parents. Foot temperatures, infant and environmental variables were measured.

RESULT: In all, 783 readings were obtained in 100 infants with a birth weight range 515 to 4885 g (mean 2152 g). The between-person correlation was 0.30 (P < 0.001) and the within-person correlation was 0.52 (P < 0.001). Bland-Altman plots showed wide 95% confidence intervals in the differences between MFH and DAT measurements (-0.87 to 1.16 °C). Differences were affected by infant variables measured. MFH more accurately predicted DAT measurements in smaller neonates and were less accurate in neonates requiring Bubble Continuous Positive Airway Pressure (CPAP).

CONCLUSION

MFH thermometry is not able to replace DAT temperature recording in the newborn intensive care.

Для цитирования: Robertson-Smith J, McCaffrey F T, Sayers R, Williams S, Taylor B J. A comparison of mid-forehead and axillary temperatures in newborn intensive care.. Journal of perinatology : official journal of the California Perinatal Association. 2014. DOI: 10.1038/jp.2014.148.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.