Comparison of Vyntus CPX and Vmax Encore 29N for indirect calorimetry: a randomised crossover study in participants of the German Diabetes Study with type 2 diabetes.

Trinks Nina, Pützer Jennifer, Sutkowski Agnieszka, Burkart Volker, Roden Michael, Kuss Oliver

СтатьяРандомизированное исследованиеСравнительное исследованиеПеревод в очереди

Аннотация

OBJECTIVES

This study compared the reliability of two metabolic cart systems, Vyntus CPX and Vmax Encore 29N, to measure whole-body energy metabolism by indirect calorimetry (IC) in individuals with type 2 diabetes (T2D).

DESIGN

Randomised, prospective, crossover study.

SETTING

Single-centre study conducted in the clinical research centre of the German Diabetes Study (GDS).

PARTICIPANTS

Five participants (3 men, 2 women, mean age 49±6 years, body mass index (BMI) 32.62±4.2 kg/m²) with T2D completed the study protocol. Eligibility requires existing participation in GDS and availability for four consecutive study days.

INTERVENTIONS: Participants underwent four IC measurements per day on four consecutive days, totalling 16 measurements per device. On each study day, two measurements with both devices, Vyntus CPX and Vmax Encore 29N, were performed in randomised order. Postcalorimetric gas calibration with normalisation was applied after each measurement.

PRIMARY OUTCOME MEASURES: Reliability of respiratory quotient (RQ) and resting energy expenditure (REE), as assessed from the coefficient of variation (CV) and 95% CIs.

RESULTS

Device comparison showed minor differences in CV (95% CI) for carbon dioxide production (VCO₂) (3.5% vs 5.3%; 95% CI -8.2% to 8.0%), oxygen consumption (VO₂) (3.4% vs 5.7%; 95% CI -9.3% to 8.2%), RQ (3.6% vs 2.3%; 95% CI -3.5% to 3.7%) and REE (3.1% vs 5.6%; 95% CI -8.4% to 7.8%). Postcalorimetric calibration did not consistently affect RQ or REE.

CONCLUSIONS

Vyntus CPX provides reliable IC measurements comparable to Vmax Encore 29N and may serve as a suitable replacement in clinical settings.

TRIAL REGISTRATION NUMBER: ClinicalTrials.gov identifier: NCT01055093.

Для цитирования: Trinks Nina, Pützer Jennifer, Sutkowski Agnieszka, Burkart Volker, Roden Michael, Kuss Oliver. Comparison of Vyntus CPX and Vmax Encore 29N for indirect calorimetry: a randomised crossover study in participants of the German Diabetes Study with type 2 diabetes.. BMJ open. 2025. DOI: 10.1136/bmjopen-2025-104761.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.