Machine Learning-Based Critical Congenital Heart Disease Screening Using Dual-Site Pulse Oximetry Measurements.

Siefkes Heather, Oliveira Luca Cerny, Koppel Robert, Hogan Whitnee, Garg Meena, Manalo Erlinda, Cresalia Nicole, Lai Zhengfeng, Tancredi Daniel, Lakshminrusimha Satyan, Chuah Chen-Nee

СтатьяМногоцентровое исследованиеПоддержка NIH: внешняя программаПеревод в очереди

Аннотация

BACKGROUND

Oxygen saturation (Spo2) screening has not led to earlier detection of critical congenital heart disease (CCHD). Adding pulse oximetry features (ie, perfusion data and radiofemoral pulse delay) may improve CCHD detection, especially coarctation of the aorta (CoA). We developed and tested a machine learning (ML) pulse oximetry algorithm to enhance CCHD detection.

METHODS AND

RESULTS

Six sites prospectively enrolled newborns with and without CCHD and recorded simultaneous pre- and postductal pulse oximetry. We focused on models at 1 versus 2 time points and with/without pulse delay for our ML algorithms. The sensitivity, specificity, and area under the receiver operating characteristic curve were compared between the Spo2-alone and ML algorithms. A total of 523 newborns were enrolled (no CHD, 317; CHD, 74; CCHD, 132, of whom 21 had isolated CoA). When applying the Spo2-alone algorithm to all patients, 26.2% of CCHD would be missed. We narrowed the sample to patients with both 2 time point measurements and pulse-delay data (no CHD, 65; CCHD, 14) to compare ML performance. Among these patients, sensitivity for CCHD detection increased with both the addition of pulse delay and a second time point. All ML models had 100% specificity. With a 2-time-points+pulse-delay model, CCHD sensitivity increased to 92.86% (P=0.25) compared with Spo2 alone (71.43%), and CoA increased to 66.67% (P=0.5) from 0. The area under the receiver operating characteristic curve for CCHD and CoA detection significantly improved (0.96 versus 0.83 for CCHD, 0.83 versus 0.48 for CoA; both P=0.03) using the 2-time-points+pulse-delay model compared with Spo2 alone.

CONCLUSIONS

ML pulse oximetry that combines oxygenation, perfusion data, and pulse delay at 2 time points may improve detection of CCHD and CoA within 48 hours after birth.

REGISTRATION: URL: https://www.clinicaltrials.gov/study/NCT04056104?term=NCT04056104&rank=1; Unique identifier: NCT04056104.

Для цитирования: Siefkes Heather, Oliveira Luca Cerny, Koppel Robert, Hogan Whitnee, Garg Meena, Manalo Erlinda, Cresalia Nicole, Lai Zhengfeng, Tancredi Daniel, Lakshminrusimha Satyan, Chuah Chen-Nee. Machine Learning-Based Critical Congenital Heart Disease Screening Using Dual-Site Pulse Oximetry Measurements.. Journal of the American Heart Association. 2024. DOI: 10.1161/jaha.123.033786.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.