DOI 10.1016/s0029-7844(99)00240-9
A bladder-neck support prosthesis for women with stress and mixed incontinence.
Аннотация
OBJECTIVE
To evaluate the safety and efficacy of a bladder-neck support prosthesis, a vaginal device designed to support the bladder neck, in women with genuine stress and mixed incontinence.
METHODS
For enrollment, incontinent women underwent a history, physical examination including cotton-swab test, urinalysis, postvoid residual urine, and multichannel urodynamic testing. Those with genuine stress incontinence or mixed incontinence and urethral hypermobility completed 7-day bladder diaries, the Incontinence Impact Questionnaire, and underwent standardized pad tests. They were fitted with a prosthesis and seen weekly to optimize fit and efficacy. At week 5, they underwent repeat evaluations with the best-fitting prosthesis in place.
RESULTS
Seventy women were enrolled and 53 completed the 1-month study (29 genuine stress incontinence, 24 mixed incontinence). The mean ages were 50.4 years for genuine stress incontinence (range 24-76) and 55.7 years for mixed incontinence (range 30-88). A statistically significant reduction in incontinence was noted on pad testing (genuine stress incontinence, mean 46.6-16.6 g; mixed incontinence, mean 31.9-6.8 g) and in the bladder diary (genuine stress incontinence, mean 28.6-7.8 losses per week; mixed incontinence, mean 30.2-15 losses per week). Quality-of-life scores improved in both groups. With the device in place, urodynamic testing indicated normalization of urethral function without evidence of outflow obstruction. Subjects found the device comfortable, easy to use, and convenient. Side effects included five urinary tract infections and 23 cases of vaginal mucosal soreness or mild irritation.
CONCLUSION
The bladder-neck support prosthesis significantly reduced involuntary urine loss in women with stress and mixed incontinence.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.