MedElevator Войти в исследования
← Поиск публикаций

DOI 10.14744/bej.2026.00087

Медицинская библиотека · Beyoglu eye journal · 2026

Comparison of Swept-Source Optical Biometry and Contact A-Scan Ultrasonography for Axial Length Measurement in Eyes with Mature Cataract.

Ceylan Ali, Simsek Emrullah, Akbas Yusuf Berk, Aydin Fahri Onur, Yildiz Burcin Kepez

СтатьяПеревод в очереди

Аннотация

OBJECTIVES

The objectives are to compare axial length (AL) measurements obtained using swept-source optical biometers and contact A-scan ultrasonography (USG) in eyes with mature cataract, and to evaluate inter-device agreement and acquisition success rates.

METHODS

This retrospective cross-sectional study included eyes with clinically defined mature cataract. AL measurements were obtained using two swept-source optical devices (Tomey - OA and Heidelberg Anterion) and contact A-scan USG (Tomey UD-800), all performed on the same day by a single examiner. Inter-device comparisons, Bland-Altman agreement analyses, and intraclass correlation coefficients (ICC) were calculated. Optical measurement failure rates were recorded.

RESULTS

AL measurements were successfully obtained with both optical biometers in 50 eyes. AL values showed strong agreement between the Tomey - OA OA-2000 and Anterion, with no significant inter-device difference. In contrast, contact A-scan USG yielded significantly shorter AL measurements compared with both optical devices (mean difference approximately 0.26-0.30 mm). Optical biometry failed to obtain AL measurements in approximately one-third of eyes with mature cataract. Inter-method reliability for AL was high (ICC >0.90).

CONCLUSION

Swept-source optical biometers provide consistent and interchangeable AL measurements in eyes with mature cataract when measurements are obtainable. However, optical acquisition failure remains common in advanced cataracts, highlighting the continued clinical relevance of A-scan USG. The systematic tendency of contact ultrasound to produce shorter AL measurements should be considered during intraocular lens power calculation.

Для цитирования: Ceylan Ali, Simsek Emrullah, Akbas Yusuf Berk, Aydin Fahri Onur, Yildiz Burcin Kepez. Comparison of Swept-Source Optical Biometry and Contact A-Scan Ultrasonography for Axial Length Measurement in Eyes with Mature Cataract.. Beyoglu eye journal. 2026. DOI: 10.14744/bej.2026.00087.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.