DOI 10.1136/thoraxjnl-2019-214177
Tablet and web-based audiometry to screen for hearing loss in adults with cystic fibrosis.
Аннотация
INTRODUCTION
Individuals with chronic lung disease (eg, cystic fibrosis (CF)) often receive antimicrobial therapy including aminoglycosides resulting in ototoxicity. Extended high-frequency audiometry has increased sensitivity for ototoxicity detection, but diagnostic audiometry in a sound-booth is costly, time-consuming and requires a trained audiologist. This cross-sectional study analysed tablet-based audiometry (Shoebox MD) performed by non-audiologists in an outpatient setting, alongside home web-based audiometry (3D Tune-In) to screen for hearing loss in adults with CF.
METHODS
Hearing was analysed in 126 CF adults using validated questionnaires, a web self-hearing test (0.5 to 4 kHz), tablet (0.25 to 12 kHz) and sound-booth audiometry (0.25 to 12 kHz). A threshold of ≥25 dB hearing loss at ≥1 audiometric frequency was considered abnormal. Demographics and mitochondrial DNA sequencing were used to analyse risk factors, and accuracy and usability of hearing tests determined.
RESULTS
Prevalence of hearing loss within any frequency band tested was 48%. Multivariate analysis showed age (OR 1.127; (95% CI: 1.07 to 1.18; p value<0.0001) per year older) and total intravenous antibiotic days over 10 years (OR 1.006; (95% CI: 1.002 to 1.010; p value=0.004) per further intravenous day) were significantly associated with increased risk of hearing loss. Tablet audiometry had good usability, was 93% sensitive, 88% specific with 94% negative predictive value to screen for hearing loss compared with web self-test audiometry and questionnaires which had poor sensitivity (17% and 13%, respectively). Intraclass correlation (ICC) of tablet versus sound-booth audiometry showed high correlation (ICC >0.9) at all frequencies ≥4 kHz.
CONCLUSIONS
Adults with CF have a high prevalence of drug-related hearing loss and tablet-based audiometry can be a practical, accurate screening tool within integrated ototoxicity monitoring programmes for early detection.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.