DOI 10.1097/bsd.0000000000000250
Outcome of Bilateral C1 Laminar Hooks Combined With C2 Pedicle Screw Fixation for the Treatment of C1-C2 Instability: A Report of 18 Cases From a Single Chinese Center.
Аннотация
STUDY
DESIGN
A retrospective technical report.
OBJECTIVE
To assess the effect of bilateral C1 laminar hooks combined with C2 pedicle screw fixation for the treatment of C1-C2 instability.
SUMMARY OF BACKGROUND DATA: Various posterior atlantoaxial fixations for C1-C2 instability have been developed. However, due to anatomic anomalies of the vertebral artery, the smallness of the pedicle, trajectories of broken screws, or a lack of surgical experience, a simple atlantoaxial fixation technique with good safety and effectiveness is urgently needed.
MATERIALS AND
METHODS
From January 2007 to September 2012, 18 patients with C1-C2 instability who underwent posterior bilateral C1 laminar hooks combined with C2 pedicle screw fixation were evaluated. Six patients had acute odontoid fractures (Anderson IIc type), 8 patients had odontoid pseudarthrosis, 3 had os odontoideum, and 1 had a traumatic rupture of the transverse ligament. The mean age at the time of surgery was 34.1 years. The clinical and radiographic analyses were performed before and after the operation and at follow-up.
RESULTS
The follow-up period was 12-78 months (with an average follow-up period of 25.6 mo). All patients were relieved of pain and their neurological symptoms were substantially improved. The postoperative JOA score improved significantly (t=-7.234, P<0.001). No neurological or vascular complications occurred in these cases. The device was placed well and had not loosened or broken and plain radiographs revealed bony fusion in 17 patients. One patient had C1 posterior arch fracture 3 weeks postoperatively and she was followed up for 18 months without revision surgery.
CONCLUSIONS
When appropriate patients were selected, bilateral C1 laminar hooks combined with C2 pedicle screw fixation can be an alternative method to treat C1-C2 instability effectively with a relatively simple procedure. Preoperative planning and evaluation were crucial for the solid atlantoaxial fusion.