BJOG : an international journal of obstetrics and gynaecology · 2002

Inflationary oscillometry provides accurate measurement of blood pressure in pre-eclampsia.

Сравнительное исследованиеОценочное исследованиеСтатьяВалидационное исследованиеНазвание и аннотация: Перевод в очереди
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Golara Moneli, Benedict Amanda, Jones Clare, Randhawa Manjit, Poston Lucilla, Shennan Andrew H

Аннотация

OBJECTIVE: To evaluate the accuracy of the OMRON-MIT inflationary oscillometric device for blood pressure measurement in pregnancy and pre-eclampsia.

DESIGN: Prospective observational study, using validation methods recommended by the British Hypertension Society (BHS) and the Association for the Advancement of Medical Instrumentation (AAMI).

SETTINGS: Antenatal clinics and ward, Guy's Hospital, London.

POPULATION: Normotensive pregnant women and those diagnosed with pre-eclampsia according to the definition of the International Society for the Study of Hypertension in Pregnancy.

METHODS: Validation according to BHS protocol.

MAIN OUTCOME MEASURES: Proportion of readings within 5, 10 and 15 mmHg (absolute differences) between the automated device and two trained, blinded observers, according to the BHS and AAMI criteria.

RESULTS: The OMRON-MIT achieved an overall BHS grade B for systolic and grade A for diastolic blood pressure measurement in both pregnancy and pre-eclampsia. The mean (SD) differences between the standard and the test device were -5 (7) mmHg for systolic and 2 (6) mmHg for diastolic blood pressure in pregnancy and -4 (6) mmHg for systolic and 2 (7) mmHg for diastolic blood pressure in pre-eclampsia. This device therefore fulfils the AAMI criteria.

CONCLUSION: The OMRON-MIT is the only automated oscillometric device that has proven to be accurate for blood pressure measurement in pre-eclampsia according to the BHS protocol in pregnancy. Inflationary oscillometry may correct the error associated with oscillometric devices in pre-eclampsia.