DOI 10.1016/j.jflm.2015.10.009

Медицинская библиотека · Journal of forensic and legal medicine · 2015

Diagnostic value of CT-localizer and axial low-dose computed tomography for the detection of drug body packing.

Aissa Joel, Kohlmeier Antonia, Rubbert Christian, Hohn Ulrich, Blondin Dirk, Schleich Christoph, Kröpil Patric, Boos Johannes, Antoch Gerald, Miese Falk

СтатьяПеревод в очереди

Аннотация

OBJECTIVES

The purpose of this study was to assess the diagnostic performance of CT-localizers in the detection of intracorporal containers.

METHODS

This study was approved by the research ethics committee of our clinic. From March 2012 to March 2013, 108 subjects were referred to our institute with suspected body packing. The CT-localizer and the axial CT-images were compared by two blinded observers retrospectively. Presence of body packs was assessed in consensus. Sensitivity and specificity, PPV and NPV of the CT-localizer were calculated.

RESULTS

Packets were detected in the CT-localizer of 19 suspects. In 28 of 108 cases packs were detected in axial CT-images. Sensitivity of CT-localizer for detection of packs was 0.68, and specificity was 1.00. There were no cases rated as false positive. The PPV was 1.0 and the NPV was 0.89. The omission of the axial CT-images would have led to a mean radiation dose reduction of 1.94 ± 0.5 mSv.

CONCLUSIONS

The value of CT-localizers lies in their high PPV. Localizers are limited by low sensitivity, compared to axial CT-images in screening of potential body packers. However, in positive cases their high PPV may possibly allow to omit the complete axial abdominal CT to achieve even lower radiation exposure.

Для цитирования: Aissa Joel, Kohlmeier Antonia, Rubbert Christian, Hohn Ulrich, Blondin Dirk, Schleich Christoph, Kröpil Patric, Boos Johannes, Antoch Gerald, Miese Falk. Diagnostic value of CT-localizer and axial low-dose computed tomography for the detection of drug body packing.. Journal of forensic and legal medicine. 2015. DOI: 10.1016/j.jflm.2015.10.009.