Tidal volume delivery during continuous chest compressions and sustained inflation.
СтатьяНазвание и аннотация: Перевод в очереди
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Solevåg A L, Lee T-F, Lu M, Schmölzer G M, Cheung P-Y
Аннотация
OBJECTIVE: To determine the distending pressure needed to achieve sufficient tidal volume (VT) delivery during continuous chest compressions (CC) superimposed by sustained inflation (SI) (CC+SI).
DESIGN: Randomised animal/manikin trial.
SETTING: University laboratory.
SUBJECTS: Cadaver piglets/manikin.
INTERVENTIONS: SI distending pressures of 5, 10, 15, 20, 25 and 30 cm H2O were delivered in random order during CC+SI for 2 min each.
MAIN OUTCOME MEASURES: VT, gas flow and airway pressure. Spearman's r for distending pressure and VT.
RESULTS: Distending pressure and VT correlated in cadaver piglets (r=0.83, p<0.001), manikin (r=0.98, p<0.001) and combined data (r=0.49, p<0.001). VT was delivered during chest recoil during CC in both models. In cadaver piglets, a distending pressure ∼25 cm H2O was needed to achieve an adequate VT.
CONCLUSIONS: Chest recoil generates VT depending on an adequate distending pressure. This has previously been demonstrated in adult animals. A pressure of ∼25 cm H2O is needed to achieve an adequate VT delivery.
DESIGN: Randomised animal/manikin trial.
SETTING: University laboratory.
SUBJECTS: Cadaver piglets/manikin.
INTERVENTIONS: SI distending pressures of 5, 10, 15, 20, 25 and 30 cm H2O were delivered in random order during CC+SI for 2 min each.
MAIN OUTCOME MEASURES: VT, gas flow and airway pressure. Spearman's r for distending pressure and VT.
RESULTS: Distending pressure and VT correlated in cadaver piglets (r=0.83, p<0.001), manikin (r=0.98, p<0.001) and combined data (r=0.49, p<0.001). VT was delivered during chest recoil during CC in both models. In cadaver piglets, a distending pressure ∼25 cm H2O was needed to achieve an adequate VT.
CONCLUSIONS: Chest recoil generates VT depending on an adequate distending pressure. This has previously been demonstrated in adult animals. A pressure of ∼25 cm H2O is needed to achieve an adequate VT delivery.