DOI 10.1136/archdischild-2016-311043

Медицинская библиотека · Archives of disease in childhood. Fetal and neonatal edition · 2016

Tidal volume delivery during continuous chest compressions and sustained inflation.

Solevåg A L, Lee T-F, Lu M, Schmölzer G M, Cheung P-Y

СтатьяПеревод в очереди

Аннотация

OBJECTIVE

To determine the distending pressure needed to achieve sufficient tidal volume (VT) delivery during continuous chest compressions (CC) superimposed by sustained inflation (SI) (CC+SI).

DESIGN

Randomised animal/manikin trial.

SETTING

University laboratory.

SUBJECTS: Cadaver piglets/manikin.

INTERVENTIONS: SI distending pressures of 5, 10, 15, 20, 25 and 30 cm H2O were delivered in random order during CC+SI for 2 min each.

MAIN OUTCOME MEASURES: VT, gas flow and airway pressure. Spearman's r for distending pressure and VT.

RESULTS

Distending pressure and VT correlated in cadaver piglets (r=0.83, p<0.001), manikin (r=0.98, p<0.001) and combined data (r=0.49, p<0.001). VT was delivered during chest recoil during CC in both models. In cadaver piglets, a distending pressure ∼25 cm H2O was needed to achieve an adequate VT.

CONCLUSIONS

Chest recoil generates VT depending on an adequate distending pressure. This has previously been demonstrated in adult animals. A pressure of ∼25 cm H2O is needed to achieve an adequate VT delivery.

Для цитирования: Solevåg A L, Lee T-F, Lu M, Schmölzer G M, Cheung P-Y. Tidal volume delivery during continuous chest compressions and sustained inflation.. Archives of disease in childhood. Fetal and neonatal edition. 2016. DOI: 10.1136/archdischild-2016-311043.