MedElevator Войти в исследования
← Поиск публикаций

DOI 10.2807/1560-7917.es.2019.24.22.1900298

Медицинская библиотека · Euro surveillance : bulletin Europeen sur les maladies transmissibles = European communicable disease bulletin · 2019

Chlamydia trachomatis samples testing falsely negative in the Aptima Combo 2 test in Finland, 2019.

Rantakokko-Jalava Kaisu, Hokynar Kati, Hieta Niina, Keskitalo Anniina, Jokela Pia, Muotiala Anna, Jokiranta T Sakari, Kuusela Rutta, Sarkkinen Hannu, Aittoniemi Janne, Vuorinen Tytti, Hakanen Antti J, Puolakkainen Mirja

СтатьяПеревод в очереди

Аннотация

Since February 2019, over 160 Chlamydia trachomatis (CT) cases testing negative or equivocal by Aptima Combo 2 (AC2) but positive by Aptima CT test run with Panther instruments occurred in Finland. The AC2 test targets chlamydial 23S rRNA while the CT test targets 16S rRNA. Sequencing of 10 strains revealed a nucleotide substitution in 23S rRNA. The significance of this for the failure of the AC2 test to detect the variant is not yet known.

Для цитирования: Rantakokko-Jalava Kaisu, Hokynar Kati, Hieta Niina, Keskitalo Anniina, Jokela Pia, Muotiala Anna, Jokiranta T Sakari, Kuusela Rutta, Sarkkinen Hannu, Aittoniemi Janne, Vuorinen Tytti, Hakanen Antti J, Puolakkainen Mirja. Chlamydia trachomatis samples testing falsely negative in the Aptima Combo 2 test in Finland, 2019.. Euro surveillance : bulletin Europeen sur les maladies transmissibles = European communicable disease bulletin. 2019. DOI: 10.2807/1560-7917.es.2019.24.22.1900298.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.