Comparison between DiaPlexQ™ STI6 and GeneFinder™ STD I/STD II multiplex Real-time PCR Kits in the detection of six sexually transmitted disease pathogens.

Huh Hee Jae, Ki Chang-Seok, Yun Sun Ae, Lee Jungsoo, Oh Gwi Young, Lee Nam-Sihk, Yoon Young Ho, Lee Nam Yong

Сравнительное исследованиеСтатьяПеревод в очереди

Аннотация

BACKGROUND

The DiaPlexQ™ STI6 Detection Kit (DiaPlexQ; Solgent Co., Ltd., Daejeon, South Korea) is a multiplex real-time PCR assay for the detection of the following sexually transmitted disease (STD) pathogens: Chlamydia trachomatis, Neisseria gonorrhoeae, Mycoplasma hominis, Trichomonas vaginalis, Ureaplasma urealyticum, and Mycoplasma genitalium. We compared the performance of the DiaPlexQ assay with the GeneFinder™ STD I (CT/NG/UU) and STD II (MG/MH/TV) Multiplex Real-time PCR Kits (GeneFinder; Infopia Co., Ltd., Anyang, South Korea).

METHODS

We evaluated the performance of the DiaPlexQ assay in comparison to that of GeneFinder using 1106 clinical specimens (542 genital swabs and 564 urine samples). The analytical performance of the DiaPlexQ assay, including the limit of detection (LOD) and analytical specificity, was evaluated using reference strains.

RESULTS

The positive percent agreement, negative percent agreement, and kappa value between the two assays were 96.6%-99.4%, 98.2%-99.8%, and 0.93%-0.99%, respectively. No cross-reactivity was observed in a collection of 41 different microorganisms and the LOD of the DiaPlexQ assay ranged from 1 to 10 copies/reaction for each microorganism.

CONCLUSION

The DiaPlexQ assay showed comparable performance to that of the GeneFinder assay so that it can be used for the screening and diagnosis of non-viral curable STD pathogens.

Для цитирования: Huh Hee Jae, Ki Chang-Seok, Yun Sun Ae, Lee Jungsoo, Oh Gwi Young, Lee Nam-Sihk, Yoon Young Ho, Lee Nam Yong. Comparison between DiaPlexQ™ STI6 and GeneFinder™ STD I/STD II multiplex Real-time PCR Kits in the detection of six sexually transmitted disease pathogens.. Journal of clinical laboratory analysis. 2018. DOI: 10.1002/jcla.22703.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.