Медицинская библиотека · Journal of the American Veterinary Medical Association · 2002

Evaluation of an oscillometric blood pressure monitor for use in anesthetized cats.

Pedersen Karen M, Butler Michael A, Ersbøll Annette K, Pedersen Henrik D

Оценочное исследованиеСтатьяПоддержка исследования: вне правительства СШАПеревод в очереди

Аннотация

OBJECTIVE

To determine accuracy of an oscillometric blood pressure monitor used over a wide range of pressures in anesthetized cats.

DESIGN

Prospective study.

ANIMALS: 6 healthy cats.

PROCEDURE: 4 female cats and 2 male cats that weighed 2.7 to 4.5 kg (5.9 to 9.9 lb) and were 2 to 8 years old were anesthetized. Blood pressure was measured directly with an arterial catheter placed in the right femoral artery and indirectly from the left antebrachium by use of an oscillometric monitor. A series of diastolic arterial pressure (DAP), mean arterial pressure (MAP), and systolic arterial pressure (SAP) measurements were obtained during hypotension, normotension, and hypertension. Values obtained indirectly and directly were compared.

RESULTS

The oscillometric monitor was accurate for DAP and MAP throughout the entire pressure range and met the standards of the Association for the Advancement of Medical Instrumentation (mean +/- SD difference from values obtained directly, < or = 5 +/- 8 mm Hg). The SAP was increasingly underestimated with increasing overall pressure; mean differences from direct measurements were -5.2, -12.1, and -17.7 mm Hg during hypo-, normo-, and hypertension, respectively. Standard deviations for SAP were all < or = 8 mm Hg. The monitor gave readings during all attempts. The direct blood pressure recording system appeared to perform well with neither under- nor overdamping.

CONCLUSIONS AND CLINICAL RELEVANCE: Except for a minor underestimation of SAP during normo- and hypertension, the oscillometric monitor yielded reliable and easily obtainable blood pressure measurements in anesthetized cats.

Для цитирования: Pedersen Karen M, Butler Michael A, Ersbøll Annette K, Pedersen Henrik D. Evaluation of an oscillometric blood pressure monitor for use in anesthetized cats.. Journal of the American Veterinary Medical Association. 2002. DOI: 10.2460/javma.2002.221.646.