MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1097/scs.0000000000002754

Медицинская библиотека · The Journal of craniofacial surgery · 2016

Comparison of Different Fixation Types Used in Unilateral Mandibular Condylar Fractures: An In Vivo Study With New Biomechanical Model.

Çimen Emre, Önder Mustafa Ercüment, Cambazoğlu Mine, Birant Emir

Сравнительное исследованиеСтатьяПеревод в очереди

Аннотация

INTRODUCTION

The aim of this in vivo study is to compare the single-titanium, double-titanium mini plate, and single resorbable plate systems used in internal rigid fixation of the unilateral mandibular condylar fractures on new design biomechanical model.

METHODS

Thirty synthetic polyurethane models were used for biomechanical testing. Fracture lines were created for each model. Fragments were fixed with single-titanium plates in Group A (n = 10), double-titanium plates in Group B (n = 10), and single biodegradable plate (PPLA) in Group C (n = 10). Masticatory forces were applied to the models and the biomechanical properties of the titanium plate and screws, resorbable plate, and screws were evaluated.

RESULTS

The average failure force for Group A, Group B, and Group C is 199, 324, 177N and the average bone displacement for Group A, Group B, Group C is 1.9, 0.3, 2.1 mm, respectively.

DISCUSSION: Double titanium plates showed the most acceptable results in the fixation of unilateral subcondylar fractures where the single titanium and biodegradable plate systems failed to provide enough stability in unilateral subcondylar fracture fixation. Biodegradable plate systems are still not an alternative in fixation of unilateral condylar fractures.

Для цитирования: Çimen Emre, Önder Mustafa Ercüment, Cambazoğlu Mine, Birant Emir. Comparison of Different Fixation Types Used in Unilateral Mandibular Condylar Fractures: An In Vivo Study With New Biomechanical Model.. The Journal of craniofacial surgery. 2016. DOI: 10.1097/scs.0000000000002754.