An In Vitro Study of Magnetic Field Interference with an Electronic Shunt Programmer.

Pajer Hengameh B, Carlson Andrew P, Botros James A, Spader Heather S

СтатьяПеревод в очереди

Аннотация

BACKGROUND

Programmable shunts play a valuable role in the treatment of hydrocephalus. However, the use of magnets in programming these valves has caused obstacles in today's magnetized world. Previous studies have reported problems with magnetic toys and electronics unintentionally reprogramming shunt valves. This study investigated how an Apple Watch can interfere with the Codman CERTAS Plus electronic programmer.

METHODS

In this in vitro study, we tested the magnetic field emitted by Apple Watch Series 3, 4, 5, and 6 using an electromagnetic field tester at distances of 0-50 mm. We conducted 20 trials of shunt programming and shunt setting reading with and without each watch.

RESULTS

All 4 watches generated significant magnetic fields. Maximum magnetic fields were Series 3, 165.73 millitesla (mT); Series 4, 144.91 mT; Series 5, 131.09 mT; and Series 6, 130.68 mT. All 4 watches interfered with the programmer's ability to correctly read and program the valve setting.

CONCLUSIONS

The Codman CERTAS Plus electronic programmer detects the magnetic field emitted from an Apple Watch and mistakes it for the valve, rendering programming difficult. These smartwatches and similar electronic devices should be kept away from the programmer and not worn by health care providers to avoid inappropriate readings and setting changes.

Для цитирования: Pajer Hengameh B, Carlson Andrew P, Botros James A, Spader Heather S. An In Vitro Study of Magnetic Field Interference with an Electronic Shunt Programmer.. World neurosurgery. 2022. DOI: 10.1016/j.wneu.2022.07.063.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.