Comparison of fluorescence intensity of protoporphyrin IX as observed on the screen of different cystoscopic systems.

Fukuhara Hideo, Nishimura Takahiro, Shimojo Yu, Inoue Keiji

СтатьяСравнительное исследованиеПеревод в очереди

Аннотация

BACKGROUND

Photodynamic diagnosis-assisted transurethral resection of bladder tumor (PDD-TURBT) is an important and useful intervention used for removing non-invasive bladder cancer (NMIBC). Several PDD devices can be used in clinical practice, but few reports have compared them.In this study, we examined the differences in detected fluorescence intensity for each PDD device.

METHODS

The PDD device was used from OLYMPUS, Richard Wolf, and Karl Storz. Light power and illuminance from endoscope were measured. For each PDD device, fluorescence intensity was measured at 3 cm and 8 cm distance from the protoporphyrin IX (PpIX) phantom to the cystoscope.

RESULTS

The variation of excitation light power with observation distance were 35.0 %, 27.0 %, and 28.2 % for OLYMPUS, Richard Wolf, and Karl Storz, respectively. The variation of fluorescence intensity with observation distance were 132.8 %, 120.5 %, and 49.8 % for OLYMPUS, Richard Wolf, and Karl Storz, respectively.

CONCLUSIONS

PDD devices used in clinical practice show both some performance differences between the PDD systems themselves, and also differences in observed fluorescence intensity.

Для цитирования: Fukuhara Hideo, Nishimura Takahiro, Shimojo Yu, Inoue Keiji. Comparison of fluorescence intensity of protoporphyrin IX as observed on the screen of different cystoscopic systems.. Photodiagnosis and photodynamic therapy. 2024. DOI: 10.1016/j.pdpdt.2024.104425.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.