Physica medica : PM : an international journal devoted to the applications of physics to medicine and biology : official journal of the Italian Association of Biomedical Physics (AIFB) · 2017
EPID-based in vivo dosimetry for stereotactic body radiotherapy of non-small cell lung tumors: Initial clinical experience.
СтатьяНазвание и аннотация: Перевод в очереди
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Consorti R, Fidanzio A, Brainovich V, Mangiacotti F, De Spirito M, Mirri M A, Petrucci A
Аннотация
PURPOSE: EPID-based in vivo dosimetry (IVD) has been implemented for stereotactic body radiotherapy treatments of non-small cell lung cancer to check both isocenter dose and the treatment reproducibility comparing EPID portal images.
METHODS: 15 patients with lung tumors of small dimensions and treated with volumetric modulated arc therapy were enrolled for this initial experience. IVD tests supplied ratios R between in vivo reconstructed and planned isocenter doses. Moreover a γ-like analysis between daily EPID portal images and a reference one, in terms of percentage of points with γ-value smaller than 1, Pγ<1, and mean γ-values, γmean, using a local 3%-3mm criteria, was adopted to check the treatment reproducibility. Tolerance levels of 5% for R ratio, Pγ<1 higher than 90% and γmean lower than 0.67 were adopted.
RESULTS: A total of 160 EPID images, two images for each therapy session, were acquired during the treatment of the 15 patients. The overall mean of the R ratios was equal to 1.005±0.014 (1 SD), with 96.9% of tests within±5%. The 2D image γ-like analysis showed an overall γmean of 0.39±0.12 with 96.1% of tests within the tolerance level, and an average Pγ<1 value equal to 96.4±3.6% with 95.4% of tests with Pγ<1>90%. Paradigmatic discrepancies were observed in three patients: a set-up error and a patient morphological change were identified thanks to CBCT image analysis whereas the third discrepancy was not fully justified.
CONCLUSIONS: This procedure can provide improved patient safety as well as a first step to integrate IVD and CBCT dose recalculation.
METHODS: 15 patients with lung tumors of small dimensions and treated with volumetric modulated arc therapy were enrolled for this initial experience. IVD tests supplied ratios R between in vivo reconstructed and planned isocenter doses. Moreover a γ-like analysis between daily EPID portal images and a reference one, in terms of percentage of points with γ-value smaller than 1, Pγ<1, and mean γ-values, γmean, using a local 3%-3mm criteria, was adopted to check the treatment reproducibility. Tolerance levels of 5% for R ratio, Pγ<1 higher than 90% and γmean lower than 0.67 were adopted.
RESULTS: A total of 160 EPID images, two images for each therapy session, were acquired during the treatment of the 15 patients. The overall mean of the R ratios was equal to 1.005±0.014 (1 SD), with 96.9% of tests within±5%. The 2D image γ-like analysis showed an overall γmean of 0.39±0.12 with 96.1% of tests within the tolerance level, and an average Pγ<1 value equal to 96.4±3.6% with 95.4% of tests with Pγ<1>90%. Paradigmatic discrepancies were observed in three patients: a set-up error and a patient morphological change were identified thanks to CBCT image analysis whereas the third discrepancy was not fully justified.
CONCLUSIONS: This procedure can provide improved patient safety as well as a first step to integrate IVD and CBCT dose recalculation.
Предметные рубрики MeSH
Carcinoma, Non-Small-Cell LungCone-Beam Computed TomographyFour-Dimensional Computed TomographyHumansIn Vivo DosimetryLungLung NeoplasmsParticle AcceleratorsPatient SafetyRadiation InjuriesRadiotherapy DosageRadiotherapy Planning, Computer-AssistedRadiotherapy, Image-GuidedRadiotherapy, Intensity-ModulatedRespirationTumor Burden