TERTmonitor Efficacy and Performance in Detecting Mutations by Droplet Digital PCR.
Аннотация
BACKGROUND
The screening of TERT promoter (TERTp) mutations is essential in cancer research and diagnostics, due to its prevalence in tumours associated with low self-renewal rates. TERTmonitor is a diagnosis kit primarily designed for real-time qPCR qualitative detection of -124C>T and -146C>T TERTp mutations, which are highly prevalent in several malignancies, particularly in bladder carcinoma.
OBJECTIVE
This study aims to investigate TERTmonitor performance in droplet digital PCR (ddPCR) in urine samples from bladder cancer patients.
METHODS
A total of 45 urine samples were examined by real-time qPCR and ddPCR techniques, and their performances were compared.
RESULTS
TERTmonitor had similar performance in both real-time qPCR and ddPCR platforms. Specifically, the methods exhibited a concordance rate of 95.45% and 90% for -124C>T and -146C>T mutations, respectively. Importantly, an enhanced sensitivity in certain scenarios was exhibited by ddPCR when compared to real-time qPCR, detecting mutations that the latter failed to identify in approximately 4.55% and 10% of the samples for -124C>T and -146C>T mutations, respectively. This enhanced sensitivity of ddPCR was particularly evident in samples with low-frequency mutations.
CONCLUSIONS
The findings highlight the usefulness of TERTmonitor for cancer surveillance either in real-time qPCR or ddPCR platforms.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.