Journal of neurotrauma · 2002
Fiberoptic intraparenchymal brain pressure monitoring with the Camino V420 monitor: reflections on our experience in 163 severely head-injured patients.
СтатьяНазвание и аннотация: Перевод в очереди
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Poca Maria-Antonia, Sahuquillo Juan, Arribas Mercedes, Báguena Marcelino, Amorós Sonia, Rubio Enrique
Аннотация
To assess the safety and accuracy of the Camino intraparenchymal sensor, we prospectively evaluated hemorrhagic complications, zero-drift, infection, and system malfunction in 163 patients monitored after a severe head injury. Mean duration of intracranial pressure (ICP) monitoring was 5 +/- 2.2 days (range: 12 h to 11 days). Of the 141 patients with a control CT scan, four showed a 1-2-cc collection of blood at the catheter's end. When removed, the sensors underread the true ICP value (negative zero-drift) in 80 of the 126 sensors evaluated (63.5%). Fourteen sensors showed no zero-drift, and 32 sensors overread the true ICP value (positive zero-drift) (median: -1 mm Hg; interquartile range: -4 to +1 mm Hg). No significant relationship was found between zero-drift, the surgeon who implanted the sensor, intracranial hypertension, or duration of ICP monitoring. No clinical infections could be attributed to the devices. Sixteen patients (9.8%) required more than one ICP sensor due to malfunctioning of the system. In conclusion, continuous ICP monitoring using the Camino intraparenchymal sensor has a low complication rate. However, this sensor may underread the real ICP values in a high number of patients. The lack of correlation between duration of ICP monitoring and zero-drift suggests that, contrary to the recommendations of other reports, the intraparenchymatous Camino sensor can provide reliable readings after the fifth day of use.