Catheterization and cardiovascular interventions : official journal of the Society for Cardiac Angiography & Interventions · 2008
Prevention and management of transcatheter balloon-expandable aortic valve malposition.
СтатьяМногоцентровое исследованиеНазвание и аннотация: Перевод в очереди
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Al Ali Abdullah M, Altwegg Lukas, Horlick Eric M, Feindel Christopher, Thompson Christopher R, Cheung Anson, Carere Ronald G, Humphries Karin, Ye Jian, Masson Jean-Bernard, Webb John G
Аннотация
OBJECTIVES: Early clinical outcomes in selected high-risk patients undergoing catheter-based aortic valve replacement (AVR) compare favorably with conventional surgical AVR. Improved understanding of the mechanisms of success and failure of transcatheter AVR will likely improve outcomes further. To this end, we examined our experience during the developmental phases of transcatheter AVR and describe the causes and management of prosthetic valve malposition.
METHODS: Transcatheter balloon-expandable AVR was performed in 170 patients at two centers. Malposition was defined as prosthetic valve implantation in a location other than within the native valve. Patients were prospectively identified and followed as part of an ongoing database.
RESULTS: Valve malposition occurred in 9 of 170 patients (5.3%). Final position was supravalvular in eight of nine cases. In all cases, embolization to the ascending aorta occurred within a few cardiac cycles following deployment. Importantly, late embolization was not observed. In most cases, the prosthesis was uneventfully repositioned in the more distal aorta. Positioning was subvalvular in one patient (0.6%), resulting in a severe regurgitation with residual native valve stenosis. Implantation of a second transcatheter valve was attempted in six patients and was successful in all. Conventional AVR was performed in two patients, with early mortality in one. At late follow-up (mean 412 days), seven of nine patients remain alive (78%) with a functioning prosthesis and relief of aortic stenosis.
CONCLUSIONS: Malposition of current balloon-expandable aortic valves is a largely preventable complication. An improved understanding of the procedure will likely minimize this possibility and mitigate the consequences should malposition occur.
METHODS: Transcatheter balloon-expandable AVR was performed in 170 patients at two centers. Malposition was defined as prosthetic valve implantation in a location other than within the native valve. Patients were prospectively identified and followed as part of an ongoing database.
RESULTS: Valve malposition occurred in 9 of 170 patients (5.3%). Final position was supravalvular in eight of nine cases. In all cases, embolization to the ascending aorta occurred within a few cardiac cycles following deployment. Importantly, late embolization was not observed. In most cases, the prosthesis was uneventfully repositioned in the more distal aorta. Positioning was subvalvular in one patient (0.6%), resulting in a severe regurgitation with residual native valve stenosis. Implantation of a second transcatheter valve was attempted in six patients and was successful in all. Conventional AVR was performed in two patients, with early mortality in one. At late follow-up (mean 412 days), seven of nine patients remain alive (78%) with a functioning prosthesis and relief of aortic stenosis.
CONCLUSIONS: Malposition of current balloon-expandable aortic valves is a largely preventable complication. An improved understanding of the procedure will likely minimize this possibility and mitigate the consequences should malposition occur.
Предметные рубрики MeSH
AgedAged, 80 and overAortic ValveAortic Valve StenosisCanadaCardiac CatheterizationCatheterizationDevice RemovalForeign-Body MigrationHeart Valve ProsthesisHeart Valve Prosthesis ImplantationHumansProspective StudiesProsthesis DesignProsthesis FailureRadiography, InterventionalTime FactorsTreatment Outcome