Closure of Long-Tunnel PFOs With the Coherex Flatstent EF - A Tailored Approach.
Аннотация
AIMS: Despite rapid progress in device technologies for patent foramen ovale (PFO) closure over the past decade, long-tunnel anatomies still constitute a challenge. The present study investigated the performance of a novel in-tunnel device (Flatstent EF; Coherex Medical) in long-tunnel PFOs.
METHODS AND
RESULTS
Three different umbrella devices (n = 61) and the Coherex Flatstent (n = 27) were used for PFO closure. The Flatstent was the preferred device in long-tunnel anatomies. Seven patients with long PFO tunnels underwent "detunnelization" by stepwise inflation of a low-pressure balloon followed by implantation of an umbrella device. Complete occlusion or trivial residual shunting ("clinical" occlusion) was achieved in 93% of the Flatstent and 92% of the umbrella device procedures (P=.92). Device performance in long-tunnel anatomies was in favor of the Flatstent (n = 24) compared with conventional occluders (n = 7), with "clinical" occlusion of 96% vs 86% (P=.24) and procedure time of 44 ± 16 minutes vs 59 ± 21 minutes (P=.04). Furthermore, postprocedural arrhythmias were significantly less frequent after Flatstent implantations (0.0% vs 9.1%; P=.03).
CONCLUSION
In long-tunnel PFOs, the Flatstent device was quicker to deploy, was at least as equally efficacious as umbrella devices, and reduced the incidence of symptomatic arrhythmias following PFO closure.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.