MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1097/xcs.0000000000001012

Медицинская библиотека · Journal of the American College of Surgeons · 2024

Clinical Validation of Computer-Aided Diagnosis Software for Preventing Retained Surgical Sponges.

Kurisaki Ken, Soyama Akihiko, Hamauzu Shin, Yamada Masahiko, Yamaguchi Shun, Matsuguma Kunihito, Kerkhof Enzo, Fukuda Toru, Toya Ryo, Eguchi Susumu

СтатьяПеревод в очереди

Аннотация

BACKGROUND

We previously reported the successful development of a computer-aided diagnosis (CAD) system for preventing retained surgical sponges with deep learning using training data, including composite and simulated radiographs. In this study, we evaluated the efficacy of the CAD system in a clinical setting.

STUDY

DESIGN

A total of 1,053 postoperative radiographs obtained from patients 20 years of age or older who underwent surgery were evaluated. We implemented a foreign object detection application software on the portable radiographic device used in the operating room to detect retained surgical sponges. The results of the CAD system diagnosis were prospectively collected.

RESULTS

Among the 1,053 images, the CAD system detected possible retained surgical items in 150 images. Specificity was 85.8%, which is similar to the data obtained during the development of the software.

CONCLUSIONS

The validation of a CAD system using deep learning in a clinical setting showed similar efficacy as during the development of the system. These results suggest that the CAD system can contribute to the establishment of a more effective protocol than the current standard practice for preventing the retention of surgical items.

Для цитирования: Kurisaki Ken, Soyama Akihiko, Hamauzu Shin, Yamada Masahiko, Yamaguchi Shun, Matsuguma Kunihito, Kerkhof Enzo, Fukuda Toru, Toya Ryo, Eguchi Susumu. Clinical Validation of Computer-Aided Diagnosis Software for Preventing Retained Surgical Sponges.. Journal of the American College of Surgeons. 2024. DOI: 10.1097/xcs.0000000000001012.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.