Evaluation of Assays to Measure Aminoglycosides in Serum: Comparison of Accuracy and Precision Based on External Quality Assessment.

Li Mengya, Ma Lingyun, Chen Chaoyang, Lei Lili, Zhai Xingyu, Luo Daohuang, Zhou Ying, Cui Yimin

СтатьяПеревод в очереди

Аннотация

BACKGROUND

Aminoglycosides require highly accurate therapeutic drug monitoring owing to their narrow therapeutic windows and toxic side effects. Therapeutic drug monitoring varies in different laboratories, and this difference is mainly due to the use of different analytical techniques. This study aimed to compare the accuracy and precision of immunoassays for the measurement of gentamicin, tobramycin, and amikacin in serum.

METHODS

Human plasma samples were spiked with known concentrations of amikacin, gentamicin, and tobramycin and dispatched to laboratories worldwide. The percentage deviation and coefficient of variation were calculated to compare the accuracy and precision among immunoassays and among antibiotics.

RESULTS

We analyzed 273, 534, and 207 amikacin, gentamicin, and tobramycin measurement results, obtained satisfactory rates of 83.9%, 86.3%, and 93.7%, and coefficients of variation ranging from 1.1% to 15.6%, 2.9% to 25.2%, and 1.8% to 27.0%, respectively. The percentage deviation ranged from -7.5% to 6.6%, -20.8% to 18.7%, and -33.2% to 41.5% for amikacin, gentamicin, and tobramycin, respectively. Significant differences were observed in accuracy and precision among assays for all antibiotics.

CONCLUSIONS

This study demonstrated high variations in results obtained from antibiotic assays conducted at different laboratories worldwide.

Для цитирования: Li Mengya, Ma Lingyun, Chen Chaoyang, Lei Lili, Zhai Xingyu, Luo Daohuang, Zhou Ying, Cui Yimin. Evaluation of Assays to Measure Aminoglycosides in Serum: Comparison of Accuracy and Precision Based on External Quality Assessment.. Therapeutic drug monitoring. 2020. DOI: 10.1097/ftd.0000000000000752.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.