DOI 10.1016/j.jchromb.2019.121808

Медицинская библиотека · Journal of chromatography. B, Analytical technologies in the biomedical and life sciences · 2019

An improved extraction protocol for therapeutic dabigatran monitoring using HPLC-MS/MS.

Kozlov Alexey V, Ramenskaya Galina V, Sychev Dmitry A, Vlasov Alexander M, Makarenkova Lyubov M, Stepanova Elena S, Gegechkori Vladimir I, Agatonovic-Kustrin Snezana, Chistyakov Victor V I

СтатьяПеревод в очереди

Аннотация

A new sample extraction protocol was developed for pharmacokinetic studies of dabigatran with high-performance liquid chromatography separation - electrospray ionization time-of-flight mass spectrometry analysis. After protein precipitation with acetonitrile, free dabigatran and its metabolites are separated into water phase by water-dichloromethane liquid-liquid extraction to purify the sample from proteins and endogenous lipophilic compounds. Chromatographic separation was achieved on an Agilent Zorbax SB-CN column (150 × 4.6 mm, 5 µm)) using 0.1% aqueous solution of formic acid and acetonitrile (80:20) as the mobile phase. Agilent Zorbax SB-CN column was selected to improve sample resolution and to avoided early elution of dabigatran previously seen when using a C18 column. The extended calibration curve was constructed from 5 to 1000 ng/L while precision and accuracy were assessed at four levels across the linear dynamic ranges. Within-run precision was <5.6% and the between-run precision was <3.9%. The method accuracy ranged from 89.8% to 104.4%. The developed method was successfully applied to 30 patient samples to evaluate antithrombotic efficacy and anticoagulant activity of dabigatran following knee endoprosthesis surgery.

Для цитирования: Kozlov Alexey V, Ramenskaya Galina V, Sychev Dmitry A, Vlasov Alexander M, Makarenkova Lyubov M, Stepanova Elena S, Gegechkori Vladimir I, Agatonovic-Kustrin Snezana, Chistyakov Victor V I. An improved extraction protocol for therapeutic dabigatran monitoring using HPLC-MS/MS.. Journal of chromatography. B, Analytical technologies in the biomedical and life sciences. 2019. DOI: 10.1016/j.jchromb.2019.121808.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.