Venous injection of a triphasic calcium-based implant in a sheep model of pulmonary embolism demonstrates minimal acute systemic effects.

Constant Caroline, Stroncek John D, Zeiter Stephan, Arens Daniel, Nehrbass Dirk, Gehweiler Dominic, Menzel Ursula, Benneker Lorin M, Hill Ronald S, Albers Christoph E

СтатьяПоддержка исследования: вне правительства СШАПеревод в очереди

Аннотация

PURPOSE: Implant leakage is the most common complication of vertebral augmentation. Alternative injectable materials must demonstrate intravascular safety comparable to or better than polymethyl methacrylate (PMMA). This study assessed the systemic effects of a triphasic calcium-based implant or PMMA injected directly into the femoral vein in a large animal model designed to mimic severe intravascular implant leakage.

METHODS

Six skeletally mature female sheep were randomly assigned (n = 3) to either the PMMA or the triphasic implant (AGN1, composition: calcium sulfate, β-tricalcium phosphate, brushite) treatment group. Femoral veins of each sheep were directly injected with 0.5 mL of implant material to mimic leakage volumes reported during PMMA vertebroplasty. To compare acute systemic effects of the materials, cardiovascular parameters, laboratory coagulation markers, and calcium and sulfate serum levels were monitored for 60 min after implant injection. Thrombotic and embolic events were evaluated by radiologic imaging, necropsy, and histopathology.

RESULTS

Heart rate, systemic arterial blood pressure, arterial oxygenation, arterial carbon dioxide content, and coagulation markers remained within physiological range after either AGN1 or PMMA injection. No blood flow interruption in the larger pulmonary vessels was observed in either group. Lung histopathology revealed that the severity of thrombotic changes after AGN1 injection was minimal to slight, while changes after PMMA injection were minimal to massive.

CONCLUSION

Acute systemic effects of intravascular AGN1 appeared to be comparable to or less than that of intravascular PMMA. Furthermore, in this preliminary study, the severity and incidence of pulmonary histological changes were lower for AGN1 compared to PMMA.

Для цитирования: Constant Caroline, Stroncek John D, Zeiter Stephan, Arens Daniel, Nehrbass Dirk, Gehweiler Dominic, Menzel Ursula, Benneker Lorin M, Hill Ronald S, Albers Christoph E. Venous injection of a triphasic calcium-based implant in a sheep model of pulmonary embolism demonstrates minimal acute systemic effects.. European spine journal : official publication of the European Spine Society, the European Spinal Deformity Society, and the European Section of the Cervical Spine Research Society. 2022. DOI: 10.1007/s00586-022-07303-x.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.