DOI 10.1177/10807683261443193

Медицинская библиотека · Journal of ocular pharmacology and therapeutics : the official journal of the Association for Ocular Pharmacology and Therapeutics · 2026

Evaluation of Stromal Riboflavin Penetration Using Slit-Lamp-Based Digital Image Processing in an Ex Vivo Iontophoresis Model.

Coskun Alaaddin, Senturk Fatih, Aygün Berke, Alıcı Cerengül, Kaba Duygu, Eski İrem Sıla

СтатьяПеревод в очереди

Аннотация

PURPOSE: To evaluate the effect of different iontophoresis current intensities on stromal riboflavin penetration for transepithelial corneal cross-linking (CXL) and to develop a simple, accessible, and quantitative image-processing workflow for its objective assessment using digital slit-lamp biomicroscopy.

METHODS

Ex vivo bovine corneas (n = 3 per group) were allocated to five groups: untreated control, topical riboflavin, and iontophoresis at 1, 3, or 5 mA for 5 min. After treatment, cobalt-blue-excited slit-lamp images were obtained. Stromal fluorescence was quantified using a Python-OpenCV workflow that converted images to hue-saturation-value space and applied segmentation-based masking to generate a fluorescence score (FS). For validation, stromal extracts were analyzed spectrophotometrically at 370 nm to determine riboflavin concentration. Finally, the association between FS and spectrophotometric riboflavin levels was examined to evaluate the utility of slit-lamp fluorescence as a noninvasive surrogate measure.

RESULTS

Mean FS increased in a current-dependent manner. Iontophoresis enhanced stromal loading, producing mean FS values of 1.04 ± 0.17 (1 mA/5 min), 1.82 ± 0.13 (3 mA/5 min), and 2.53 ± 0.25 (5 mA/5 min), with higher currents also promoting a more homogeneous stromal distribution. Based on the linear FS-concentration relationship, these values corresponded to increasing stromal riboflavin concentrations. A strong linear correlation (R2 = 0.96) between FS and spectrophotometrically calculated riboflavin concentration demonstrated that slit-lamp-derived fluorescence accurately predicts stromal riboflavin content.

CONCLUSION

The proposed digital slit-lamp analysis provides a simple, objective, and reproducible method for quantifying stromal riboflavin fluorescence. The strong correlation between FS and spectrophotometric measurements supports its use as a noninvasive surrogate for stromal riboflavin assessment. With further in vivo validation, this approach may enable real-time monitoring and optimization of transepithelial CXL protocols.

Для цитирования: Coskun Alaaddin, Senturk Fatih, Aygün Berke, Alıcı Cerengül, Kaba Duygu, Eski İrem Sıla. Evaluation of Stromal Riboflavin Penetration Using Slit-Lamp-Based Digital Image Processing in an Ex Vivo Iontophoresis Model.. Journal of ocular pharmacology and therapeutics : the official journal of the Association for Ocular Pharmacology and Therapeutics. 2026. DOI: 10.1177/10807683261443193.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.