Predictors of Need for Permanent Pacemaker Implantation and Conduction Abnormalities With a Novel Self-expanding Transcatheter Heart Valve.

Pellegrini Costanza, Husser Oliver, Kim Won-Keun, Holzamer Andreas, Walther Thomas, Rheude Tobias, Mayr Nicola Patrick, Trenkwalder Teresa, Joner Michael, Michel Jonathan, Chaustre Fabian, Kastrati Adnan, Schunkert Heribert, Burgdorf Christof, Hilker Michael, Möllmann Helge, Hengstenberg Christian

СтатьяМногоцентровое исследованиеПеревод в очереди

Аннотация

INTRODUCTION AND

OBJECTIVES

The incidence of permanent pacemaker implantation (PPI) and new conduction abnormalities (CA) with the ACURATE neo (Symetis S.A., Eclubens, Switzerland) has not been studied in detail. We aimed to analyze their predictors, evaluating patient- and device-related factors, including implantation depth and device-to-annulus ratio (DAR).

METHODS

Two analyses of a multicenter population were performed: new PPI in pacemaker-naive patients (n = 283), and PPI/new-CA in patients without prior CA or pacemaker (n = 232).

RESULTS

A new PPI was required in 9.9% of patients, who had a higher body mass index, higher rate of right bundle branch block and bradycardia. Neither implantation depth nor DAR differed in patients with PPI compared with those without. In the multivariable analysis neither DAR (OR, 1.010; 95%CI, 0.967-1.055; P = .7) nor implantation depth (OR, 0.972; 95%CI, 0.743-1.272; P = .8) predicted PPI. Only high body mass index, bradycardia and right bundle branch block persisted as independent predictors. PPI/new-onset CA occurred in 22.8% of patients and was associated with a higher logistic EuroSCORE. Neither implantation depth nor DAR differed in patients with PPI/new-CA vs those without (7.3 ± 1.9 vs 7.1 ± 1.5mm; P = .6 and 41.0 ± 7.9 vs 42.2 ± 10.1%; P = .4). The only predictor of PPI/new-CA was a higher logistic EuroSCORE (OR, 1.039; 95%CI, [1.008-1.071]; P = .013).

CONCLUSIONS

New PPI and new-onset CA rates were low with the ACURATE neo. These were mainly influenced by patient characteristics and not by device-depending factors.

Для цитирования: Pellegrini Costanza, Husser Oliver, Kim Won-Keun, Holzamer Andreas, Walther Thomas, Rheude Tobias, Mayr Nicola Patrick, Trenkwalder Teresa, Joner Michael, Michel Jonathan, Chaustre Fabian, Kastrati Adnan, Schunkert Heribert, Burgdorf Christof, Hilker Michael, Möllmann Helge, Hengstenberg Christian. Predictors of Need for Permanent Pacemaker Implantation and Conduction Abnormalities With a Novel Self-expanding Transcatheter Heart Valve.. Revista espanola de cardiologia (English ed.). 2018. DOI: 10.1016/j.rec.2018.01.011.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.