Nickel elution properties of contemporary interatrial shunt closure devices.
Аннотация
OBJECTIVES
We sought to compare nickel elution properties of contemporary interatrial shunt closure devices in vitro.
INTRODUCTION
There are two United States Food and Drug Administration (FDA)-approved devices for percutaneous closure of secundum atrial septal defect: the Amplatzer septal occluder (ASO; St Jude Medical Corporation) and Gore Helex septal occluder (HSO; W.L. Gore & Associates). The new Gore septal occluder (GSO) device is in clinical trials. These are also used off-label for patent foramen ovale closure in highly selected patients. These devices have high nickel content. Nickel allergy is the most common reason for surgical device explantation. Nickel elution properties of contemporary devices remain unknown.
METHODS
We compared nickel elution properties of 4 devices - ASO, GSO, HSO, and sternal wire (SW) - while Dulbecco's phosphate-buffered saline (DPBS) served as control. Three samples of each device were submerged in DPBS. Nickel content was measured at 14 intervals over 90 days.
RESULTS
Nickel elution at 24 hours, compared to control (0.005 ± 0.0 mg/L), was significantly higher for ASO (2.98 ± 1.65 mg/L; P=.04) and SW (0.03 ± 0.014 mg/L; P=.03). Nickel levels at 90 days, compared to control (0.005 ± 0.0 mg/L) and adjusting for multiple comparisons, were significantly higher for ASO (19.80 ± 2.30 mg/L; P=.01) and similar for HSO (P=.34), GSO (P=.34), and SW (P=.34). ASO had significantly higher nickel elution compared to HSO, GSO, and SW (P=.01).
CONCLUSION
There is substantial variability in nickel elution; devices with less exposed nickel (HSO and GSO) have minimal elution. The safety of low nickel elution devices in patients with nickel allergy needs to be evaluated in prospective trials.