Efficacy of a low-cost bubble CPAP system in treatment of respiratory distress in a neonatal ward in Malawi.

Kawaza Kondwani, Machen Heather E, Brown Jocelyn, Mwanza Zondiwe, Iniguez Suzanne, Gest Al, Smith E O'Brian, Oden Maria, Richards-Kortum Rebecca R, Molyneux Elizabeth

СтатьяПоддержка исследования: вне правительства СШАПоддержка правительства СШАПеревод в очереди

Аннотация

BACKGROUND

Respiratory failure is a leading cause of neonatal mortality in the developing world. Bubble continuous positive airway pressure (bCPAP) is a safe, effective intervention for infants with respiratory distress and is widely used in developed countries. Because of its high cost, bCPAP is not widely utilized in low-resource settings. We evaluated the performance of a new bCPAP system to treat severe respiratory distress in a low resource setting, comparing it to nasal oxygen therapy, the current standard of care.

METHODS

We conducted a non-randomized convenience sample study to test the efficacy of a low-cost bCPAP system treating newborns with severe respiratory distress in the neonatal ward of Queen Elizabeth Central Hospital, in Blantyre, Malawi. Neonates weighing >1,000 g and presenting with severe respiratory distress who fulfilled inclusion criteria received nasal bCPAP if a device was available; if not, they received standard care. Clinical assessments were made during treatment and outcomes compared for the two groups.

FINDINGS: 87 neonates (62 bCPAP, 25 controls) were recruited. Survival rate for neonates receiving bCPAP was 71.0% (44/62) compared with 44.0% (11/25) for controls. 65.5% (19/29) of very low birth weight neonates receiving bCPAP survived to discharge compared to 15.4% (1/13) of controls. 64.6% (31/48) of neonates with respiratory distress syndrome (RDS) receiving bCPAP survived to discharge, compared to 23.5% (4/17) of controls. 61.5% (16/26) of neonates with sepsis receiving bCPAP survived to discharge, while none of the seven neonates with sepsis in the control group survived.

INTERPRETATION: Use of a low-cost bCPAP system to treat neonatal respiratory distress resulted in 27% absolute improvement in survival. The beneficial effect was greater for neonates with very low birth weight, RDS, or sepsis. Implementing appropriate bCPAP devices could reduce neonatal mortality in developing countries.

Для цитирования: Kawaza Kondwani, Machen Heather E, Brown Jocelyn, Mwanza Zondiwe, Iniguez Suzanne, Gest Al, Smith E O'Brian, Oden Maria, Richards-Kortum Rebecca R, Molyneux Elizabeth. Efficacy of a low-cost bubble CPAP system in treatment of respiratory distress in a neonatal ward in Malawi.. PloS one. 2014. DOI: 10.1371/journal.pone.0086327.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.