MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1007/s00402-012-1628-y

Медицинская библиотека · Archives of orthopaedic and trauma surgery · 2012

Distal radius fracture fixation with volar locking plates and additional bone augmentation in osteoporotic bone: a biomechanical study in a cadaveric model.

Högel F, Mair S, Eberle S, Weninger P, von Oldenburg G, Augat P

СтатьяПоддержка исследования: вне правительства СШАПеревод в очереди

Аннотация

BACKGROUND

Fractures of the distal radius represent the most common fractures in adults. Volar locked plating has become a popular method for treating these fractures, but has been subject to several shortcomings in osteoporotic bone, such as loss of reduction and screw purchase. In order to overcome these shortcomings, cement augmentation has been proposed.

METHODS

AO-type 23-A3.3 fractures were made in 8 pairs of fresh frozen osteoporotic cadaveric radial bones. All specimens were treated with volar plating, and divided into cement augmentation or non-augmentation groups (n = 8/group). Constructs were tested dynamically and load to failure, construct-stiffness, fracture gap movement and screw cutting distance were measured.

RESULTS

Cement augmentation resulted in a significant increase in cycles and load to failure, as well as construct stiffness at loads higher than 325 N. When compared to the non-augmented group, fracture gap movement decreased significantly at this load and higher, as did screw cutting distance at the holes of the ulnar column. The cycles to failure depend on the BMD in the distal region of the radius.

CONCLUSION

Cement augmentation improves biomechanical properties in volar plating of the distal radius.

Для цитирования: Högel F, Mair S, Eberle S, Weninger P, von Oldenburg G, Augat P. Distal radius fracture fixation with volar locking plates and additional bone augmentation in osteoporotic bone: a biomechanical study in a cadaveric model.. Archives of orthopaedic and trauma surgery. 2012. DOI: 10.1007/s00402-012-1628-y.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.