BACKGROUND

The purpose of this study is to compare a new temporary triple-lumen catheter (TLC) for dialysis that has a third lumen devoted to fluid and medication administration or blood sampling with a marketed dual-lumen catheter (DLC).

METHODS

Four hundred eighty-five patients referred for acute hemodialysis or apheresis were randomly assigned to either a TLC or DLC in a multicenter, prospective, randomized trial.

RESULTS

Analysis of blood flow rates was completed on 464 patients (228 patients, DLC; 236 patients, TLC) with a total of 1,681 hemodialysis (808 treatments, DLC; 873 treatments, TLC) and 82 apheresis treatments (37 treatments, DLC; 45 treatments, TLC). During hemodialysis, a median achieved flow rate (AFR) of 267 mL/min was realized for both groups (P = 0.58). During apheresis, a median AFR of 72.5 mL/min (range, 50 to 150 mL/min) was achieved in the DLC group, and 87 mL/min (range, 60 to 150 mL/min), in the TLC group (P = 0.14). Three hundred ninety-three patients (193 patients, DLC; 200 patients, TLC) had blood and catheter tip cultures performed on removal, and catheter-related bloodstream infection (CRBSI) status was determined. Thirty-one patients (7.9%) had a CRBSI: 16 patients (8.3%), DLC; and 15 patients (7.5%), TLC (P= 0.77). Incidence densities of CRBSI were 12.4/1,000 DLC-days and 10.2/1,000 TLC-days (P = 0.59). The CRBSI incidence of 18.2/1,000 catheter-days for femoral sites was significantly greater than the 7/1,000 catheter-days for jugular sites (P = 0.02) and 6.6/1,000 catheter-days for combined jugular and subclavian sites (P = 0.01). In multivariate analysis, antibiotic use was the only factor related to CRBSI (odds ratio, 0.30; 95% confidence interval, 0.12 to 0.76). There were no statistically significant differences in rates of other complications between the 2 catheters.

CONCLUSION

Results show that the new TLC is similar to the marketed DLC.

Для цитирования: Contreras Gabriel, Liu Ping-Yu, Elzinga Lawrence, Anger Michael S, Lee Jean, Robert Nurian, Chvala Robert, Mars Ronald L, Vesely Thomas, Taber Tim E, Shemin Douglas, Shafritz Randy, Pulliam Joseph. A multicenter, prospective, randomized, comparative evaluation of dual- versus triple-lumen catheters for hemodialysis and apheresis in 485 patients.. American journal of kidney diseases : the official journal of the National Kidney Foundation. 2003. DOI: 10.1016/s0272-6386(03)00657-7.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.