Validation of the oscillometric blood pressure monitor Vet20 SunTech in anesthetized healthy cats.

Cremer Jeannette, da Cunha Anderson, Aulakh Karanvir, Liu Chin-Chi, Acierno Mark J

СтатьяВалидационное исследованиеПеревод в очереди

Аннотация

OBJECTIVE

To evaluate a veterinary-specific oscillometric noninvasive blood pressure (NIBP) system according to the guidelines of the American College of Veterinary Internal Medicine (ACVIM) Consensus Statement.

STUDY

DESIGN

Prospective clinical study.

ANIMALS: A total of 33 client-owned cats (20 females and 13 males).

METHODS

Cats were premedicated with methadone (0.3 mg kg-1) and alfaxalone (2 mg kg-1) intramuscularly. After 15 minutes anesthesia was induced with isoflurane (3%) in 100% oxygen by facemask while breathing spontaneously. A 22 gauge catheter was placed in the median caudal artery and systolic (SAP), diastolic (DAP) and mean (MAP) arterial pressures were measured. NIBP measurements were collected by placing the cuff (40% of limb circumference) on the right or left antebrachium. The agreement between the two methods was evaluated with the Bland-Altman methods, and the oscillometric NIBP device was evaluated using the ACVIM guidelines for validation of devices.

RESULTS

Data from 30 of the 33 cats were analyzed. Five paired measurements were taken from each cat, totaling 150 paired measurements. Mean bias (limits of agreements) for SAP, DAP and MAP were 2.7 (-22.7 to 28.1), 0.9 (-22.3 to 24.2) and 1.3 (-20.4 to 23.0). The oscillometric NIBP passed all validation criteria, except correlation which was <0.9 for SAP, DAP and MAP.

CONCLUSIONS AND CLINICAL RELEVANCE: The Vet20 did not meet all validation criteria by the ACVIM. However, all criteria except correlation were met.

Для цитирования: Cremer Jeannette, da Cunha Anderson, Aulakh Karanvir, Liu Chin-Chi, Acierno Mark J. Validation of the oscillometric blood pressure monitor Vet20 SunTech in anesthetized healthy cats.. Veterinary anaesthesia and analgesia. 2020. DOI: 10.1016/j.vaa.2019.12.007.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.